SECTION 7—TROUBLESHOOTING GUIDE
Perfecto2
â„¢Series 30 Part No 1143482
Alarm:
Rapid
Beep....Beep...
Beep...Beep
LOW FLOW ALARM
1a. Kinked or blocked
tubing, cannula or
humidifier.
1b. Flowmeter set at 0.5
L/min for Perfecto2 5.
1a. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power Off for
60 seconds and then turn
power back ON.
1b. Reset flowmeter to
prescribed flowrate. If
prescribed flowrate is below
0.5 L/min, use the Pediatric
Flowmeter IRCPF16 (Only
with Perfecto2 5).
SYMPTOM: PROBABLE CAUSE: SOLUTION: