SERVICING / TROUBLESHOOTING

SERVICING

Exact service intervals depend on actual usage, however, as this medical device is to be relied upon for several years, we recommend that it is serviced by a qualified techni- cian at least once a year.

To assist you, we have listed possible problems in the Troubleshooting Section of this manual.

TROUBLESHOOTING

PROBLEM

POSSIBLE SOLUTIONS

Unit does not run from battery.

1.

Assure battery is installed properly.

 

2.

Remove rubber bumpers.

 

3.

Check all leads are removed from DC

 

 

inlet.

 

4.

Charge battery for 16 hours.

 

 

 

Unit does not charge and

1.

Confirm power pack is plugged into live

GREEN Charging indicator is

 

wall outlet.

NOT illuminated.

2. Have Freeway Lite serviced by your Dealer.

Unit does not run from DC inlet.

1.

Check fuse in DC cigarette lighter lead.

 

2.

Check for correct polarity (Refer to SPECI-

 

 

FICATIONS Section in this manual).

Unit runs slowly or battery us-

1.

Check filter is clean. Replace.

age time decreased.

2.

Check coiled Duratube is NOT blocked. If

 

 

blocked, replace with new one. DO NOT

 

 

immerse in water.

 

3.

Follow Discharge/Charge directions in

 

 

"Prolonging the Life of your Battery".

 

 

(Refer to MAINTENANCE Section of

 

 

this Manual.) Procedure may need to be

 

 

repeated.

Unit runs and Sidestream

1. Thoroughly clean the nebulizer by boiling

nebulizer makes poor or no

 

in water with two (2) drops of dish wash-

mist.

 

ing liquid for 5 to 10 minutes.

 

2.

DO NOT use brush.

 

3.

Replace nebulizer with a new one.

NOTE: If any of the above problems occur and the solutions do not work, please con- tact your Invacare Dealer for service.