280274_05-054:05-054 1/21/08 12:36 PM Page 15

Invacare® Patient Monitoring Systems

AB C D

E

(A) IRC700

(B) IRC725

(C) IRC735 (D) IRC775 (E) IRC600

 

 

 

 

 

Patient Monitoring Systems Specifications

 

 

 

 

 

 

 

Digit® pulse oximeter

Pulse oximeter

Printing pulse oximeter

Sleep printing pulse oximeter

Digital pulse oximeter

 

 

 

 

E (IRC600)

A (IRC700)

B (IRC725)

C (IRC735)

D (IRC775)

 

Displays, indicators & keys

 

 

 

 

 

 

SpO2

LED numeric display,

LED numeric display,

LED numeric display,

LED numeric display,

LED numeric display,

 

 

 

 

0.32" (8.1 mm) high

0.43" (10.9 mm) high

0.43" (10.9 mm) high

0.3" (7.62 mm) high

0.3" (7.62 mm) high

 

 

Pulse rate

LED numeric display,

LED numeric display,

LED numeric display,

LED numeric display,

LED numeric display,

 

 

 

 

0.32" (8.1 mm) high

0.375" (9.5 mm) high

0.375" (9.5 mm) high

0.3" (7.62 mm) high

0.3" (7.62 mm) high

 

 

Pulse strength

LED numeric display,

logarithmically scaled

logarithmically scaled

logarithmically scaled

logarithmically scaled

 

 

 

 

0.32" (8.1 mm) high

8-segment bar graph

8-segment bar graph

8-segment bar graph

8-segment bar graph

 

 

Alarm silence

n/a

n/a

n/a

LED flashes or lights to

LED flashes or lights to

 

 

 

 

 

 

 

remind user that audio

remind user that audio

 

 

 

 

 

 

 

alarms are disabled

alarms are disabled

 

 

Low battery

indicator flashes

left bar indicator lights

1 segment LED lights

LED flashes when

LED flashes when

 

 

 

 

when 30 minutes

when less than 2 hours

 

approximately 30 min.

approximately 30 min.

 

 

 

 

remain

remain

 

of battery life remains

of battery life remains

 

 

Probe

unit shuts down

if unit is not connected,

if unit is not connected,

LED flashes to alert

LED flashes to alert

 

 

 

 

after 8 seconds

system shuts off

system shuts off

operator to check

operator to check

 

 

 

 

when not in use

in 2 minutes

in 2 minutes

SpO2 probe placement

SpO2 probe placement

 

SpO2

 

 

 

 

 

 

 

Range

0-99%

0-99%

0-99%

0-99%

0-99%

 

 

Accuracy

± 2% at 70%-99%

± 2% at 70%-100%;

± 2% at 70%-90%

± 2% at 70%-99%

± 2% at 70%-99%;

 

 

 

 

 

± 3% at 50%-69%

 

 

± 3% at 50%-69%

 

 

Averaging

8-pulse beat average

8-pulse beat average

8-pulse beat average

4-pulse beat average

4-, 8- or 16- pulse

 

 

 

 

 

 

 

 

beat average

 

 

Alarm range

n/a

n/a

n/a

n/a

high – 50%-99%

 

 

 

 

 

 

 

 

and OFF (1% steps)

 

 

 

 

 

 

 

 

low – OFF and

 

 

 

 

 

 

 

 

50%-98% (1% steps)

 

Pulse rate

 

 

 

 

 

 

 

Range

30-254 bpm

30-254 bpm

30-254 bpm

30-254 bpm

30-254 bpm

 

 

Accuracy

± 2%

± 2%

± 2%

± 2%

± 2%

 

 

Averaging

8-second average

8-second average

8-second average

8-second average

8- or 16-second average

 

 

Alarm range

n/a

n/a

n/a

n/a

high – 5-254 bpm

 

 

 

 

 

 

 

 

and OFF (1 bpm steps)

 

 

 

 

 

 

 

 

low – OFF and 5-254 bpm

 

 

 

 

 

 

 

 

(1 bpm steps)

 

Memory

n/a

99 spot-checks

99 spot-checks

99 patients or

99 patients or

 

 

 

 

 

 

 

102 hours of run time

102 hours of run time

 

Dimensions

1.7" h x 2.25" w x 1.5" d

6.30" h x 3.25" w x 1.25" d

6.60" h x 2.75" w x 1.43" d

6.60" h x 2.75" w x 1.43" d

6.60" h x 2.75" w x 1.43" d

 

 

 

 

4.3 cm x 5.7 cm x 3.8 cm

16.0 cm x 8.3 cm x 3.2 cm

16.8 cm x 7.0 cm x 3.6 cm

16.8 cm x 7.0 cm x 3.6 cm

16.8 cm x 7.0 cm x 3.6 cm

 

Weight

3 oz. (.09 kg)

16.5 oz. (0.5 kg)

13 oz. (0.4 kg)

13 oz. (0.4 kg)

16 oz. (0.5 kg)

 

 

 

 

with batteries

with batteries

with batteries

with batteries

with batteries

 

Auxiliary serial printer/computer output

 

 

 

 

 

 

 

 

n/a

SpO2 and pulse-rate

SpO2 and pulse-rate values

n/a

SpO2, pulse-rate and

 

 

 

 

 

values output every 5

output every 5 seconds

 

pulse strength values

 

 

 

 

 

seconds to optional

 

 

output every 5 seconds

 

 

 

 

 

 

 

 

to optional printer; spot-

 

 

 

 

 

 

 

 

check and graphic trends

 

 

 

 

 

 

 

 

of stored data

 

Power requirements

 

 

 

 

 

 

 

 

Power source

2 “AAA” cell alkaline

3 “C” cell alkaline

4 “AA” cell alkaline

4 “AA” cell alkaline

6 “AA” cell alkaline

 

 

 

 

batteries

or NiCad® batteries

batteries

batteries

batteries

 

 

 

Battery life

approximately

approximately

14 hours with continuous

2 overnight sleep

2 overnight sleep

 

 

 

 

1,400 spot-checks

1,500 spot-checks

printing

screening cycles

screening cycles

 

Safety approvals

UL & CSA listed

UL & CSA listed

UL & CSA listed

UL & CSA listed

UL and CSA listed

 

Limited warranty

1 year

2 years

2 years

2 years

2 years

Invacare® Patient Monitoring Systems

15

Page 15
Image 15
Invacare XPO2 manual IRC735 D IRC775 E IRC600, Patient Monitoring Systems Specifications