280274_05-054:05-054 1/21/08 12:36 PM Page 6

Invacare® Cylinders and Oxygen Delivery Device

Cylinder Specifications

Model Number

HF2PCM2KIT

HF2PCL4

HF2PCL6

HF2PC9

Height

 

8.25"

9"

11"

14"

 

 

 

 

 

 

Weight

 

2 lb.

3.1 lb.

3.5 lb.

4.8 lb

 

 

 

 

 

 

Diameter

 

3"

4.38"

4.38"

4.38"

Capacity

 

57 Liters

113 Liters

170 Liters

255 Liters

 

 

 

 

Conserved Durations (Hrs) @ 20 bpm

 

 

 

 

 

 

 

 

At setting of 1

3.4 Hrs

6.7 Hrs

9.8 Hrs

15.2 Hrs

 

 

 

 

 

 

At setting of

2

2.2 Hrs

4.3 Hrs

6.3 Hrs

10.5 Hrs

 

 

 

 

 

 

At setting of

3

1.6 Hrs

3.2 Hrs

4.7 Hrs

7.3 Hrs

 

 

 

 

 

 

At setting of

4

1.3 Hrs

2.6 Hrs

3.8 Hrs

5.9 Hrs

 

 

 

 

 

 

At setting of

5

1.1 Hrs

2.1 Hrs

3.1 Hrs

4.8 Hrs

 

 

 

 

 

 

Fill Time

 

24 Min

55 Min

75 Min

125 Min

Flow Settings

 

1-5 Pulse

1-5 Pulse

1-5 Pulse

1-5 Pulse

 

 

or 2 lpm Cont.

or 2 lpm Cont.

or 2 lpm Cont.

or 2 lpm Cont.

Invacare® Cylinders and

Cylinder Specifications

Model Number

HF2POST6

HF2POST9

HF2POSTD

IOH2M6

IOH2M9

IOH2MD

 

 

 

 

 

 

 

Height

12"

15"

21"

15.25"

14.25"

19.75"

 

 

 

 

 

 

 

Weight

4.2 lb.

5.4 lb.

6.5 lb.

3.5 lb.

4.8 lb.

7 lb.

 

 

 

 

 

 

 

Diameter

4.38"

4.38"

4.38"

3"

4.38"

4.38"

 

 

 

 

 

 

 

Capacity

170 Liters

255 Liters

422 Liters

148 Liters

248 Liters

422 Liters

 

 

 

 

 

 

 

 

Duration @ 2 lpm flow

Varies

Varies

Varies

1

Hr

1.9 Hrs

3.5 Hrs

 

 

 

 

 

 

 

 

Fill Time

75 Min

125 Min

3.5 Hrs

1

Hr

1.9 Hrs

3.5 Hrs

 

 

 

 

 

 

 

 

Flow Settings

N/A

N/A

N/A

0

- 6 lpm

0 - 6 lpm

0 - 6 lpm

 

 

 

 

 

 

 

 

Oxygen Delivery Device

Invacare® Demand Oxygen Delivery Device

Model no. IDD20EX

Invacare’s Venture® Demand Oxygen Delivery Device (DODD) is designed to give patients the freedom to do more of what they want to do. When used with smaller cylinders, it is lightweight, portable and extremely easy to carry. Extended oxygen usage time makes it perfect for out-of-home use.

Through a wide range of respiratory rates (10-40 bpm), the Venture DODD is able to deliver precise amounts of oxygen at the very beginning of each breath in which delivery duration is dependent on respiratory rate (variable-pulse technology). And its alarms are clearly identified, making solutions simple and quick.

Features

Variable demand will provide up to 4:1 conservation ratio

Internal clinicians switch will provide up to 6:1 conservation ratio

Simple operation using only 2 buttons (on and off)

Uses one D-size disposable battery

Switches easily from continuous to pulse modes; no disconnection required

Includes small and lightweight pack systems for easy carrying

Provides easy-to-use regulator for setting flow

Specifications

Battery usage time

approx. 80 hours at 20 bpm

Flow rate range

0-6 lpm

Oxygen source pressure

20 psig (± 3 psi)

Regulator

 

Flow settings

0-6 lpm

Regulated pressure

20 psig

Length

5.4"

Thickness

1.6"

Width

3.6"

Weight

16 oz.

Model Selection Guide

Model no.

Units/Case

Limited warranty

Description

IDD20EX

1

3 years

20 psi DODD with connecting tube

IRC2121

1

6 years

20 psi piston regulator, 0-6 lpm

2002919

1

1 year

DODD pouch (attaches on cylinders)

M6/M9 BAG

1

1 year

Shoulder bag, universal (M6/M9)

6

Page 6
Image 6
Invacare XPO2 manual Invacare Cylinders and Oxygen Delivery Device, Invacare Demand Oxygen Delivery Device, Specifications