5.2Commissioning the fluorescence system

After switching on the high-pressure mer- cury lamp, you must adjust the burner immedi- ately (see User Manual M2-216-1). Do not switch on the supply unit until you can safely operate the Leica MZ16 F.

Follow all safety instructions when handling mercury lamps and the lamp housing!

￿Switch on the supply unit (see User Manual M2-216-1).

Wait for 2 to 3 minutes.

￿Adjust the burner in accordance with User Manual M2-216-1.

For observation with fluorescence light, always position the aperture protection correctly to protect the observer (page 16).

2

1

Fig. 9 Filter changer, equipped with filter holders

1Filter sets

2Filter slide for filter holder with individual filters

5.3 Rapid filter changer

The rapid filter changer must be equipped with four filter holders at all times.

If you are using fewer than four filter hold- ers with fluorescence filters, always insert the included simple filter holders into the available positions. Otherwise there is a risk of danger to the eyes from direct UV radiation (see page 16).

If at all possible, avoid touching the filters in the filter sets in order to avoid fingerprints.

If you touch them, clean the filters using a soft, lint-free cloth and pure alcohol.

￿Insert a filter set into the rapid filter changer such that the label (e.g. GFP) is upright and leg- ible on the right side.

Ensure that the contour of the filter set matches the contour of the microscope.

￿Rotate the filter set by hand until it latches into place and insert the next filter set.

Insert a total of four filter sets or the included sim- ple filter holders into the empty positions.

20

Leica MZ16 F – Functions

Page 20