?WHAT DISPLAY SYMBOLS MEAN

Display

Condition/Cause

Recommended

Symbol

 

Action

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The symbol shown while measure-

Measurement is in progress, remain

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ment is in progress. It blinks while

as still as possible.

 

 

 

 

 

 

 

 

 

detecting the pulse.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

An irregular heartbeat or body move-

Take measurement again and con-

 

 

 

 

 

 

 

 

 

ment may have occurred. Refer to

sult with your physician.

 

 

 

 

 

 

 

 

 

page E-9 for more information on

 

 

 

 

 

 

 

 

 

 

irregular heartbeats.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Previous measurement stored in

 

 

 

 

 

 

 

 

 

 

MEMORY.

 

 

 

 

 

 

 

 

 

 

The battery power indicator during

Replace all batteries with new ones

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Battery Full

measurement.

when the symbol blinks. Alkaline

 

 

Battery Low

The batteries are low when it blinks.

batteries recommended.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Unstable blood pressure due to

Try the measurement again. Remain

 

 

 

 

 

 

 

 

 

excessive body movement.

very still during the measurement.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The systolic and diastolic values are

Fasten the cuff correctly, and try the

 

 

 

 

 

 

 

 

 

within 10 mmHg of each other.

measurement again.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The pressure value did not increase

Check for air leaks along the tube

 

 

 

 

 

 

 

 

 

during inflation.

and around the

 

 

 

 

 

 

 

 

 

 

air socket.

 

 

 

 

 

 

 

 

 

The cuff is not fastened correctly.

Refasten the cuff and retake meas-

 

 

 

 

 

 

 

 

 

 

urement.

 

 

 

 

 

 

 

 

 

There is an air leak in the cuff or

Make sure tube is properly connect-

 

 

 

 

 

 

 

 

 

monitor.

ed to cuff and monitor.

 

 

 

 

 

 

 

 

 

The pulse is not detected correctly.

Try the measurement again. Remain

PULSE DISPLAY ERROR

 

very still during the measurement.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Cuff Inflation Meter

Measurement is in progress, remain

 

 

 

 

 

 

 

 

 

 

as still as possible.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pressure Rating Indicator™

Refer to the section “About Pressure

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rating Indicator” for further expla-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

nation.

 

 

 

 

 

 

 

 

 

Appears on display prior to and dur-

Press exhaust button to release air.

 

 

 

 

 

 

 

 

 

ing measurement. Blinks if there is air

 

 

 

 

 

 

 

 

 

 

in cuff at power up.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Cuff needs to be inflated for proper

Inflate cuff 30 to 40 mmHg above

 

 

 

 

 

 

 

 

 

measurement.

your normal systolic pressure.

 

 

 

 

 

 

 

 

 

 

 

E-2 ?