Lumiscope 1133 instruction manual Automatic Inflation, Rapid Deflation During Measurement

Models: 1133

1 48
Download 48 pages 15 Kb
Page 13
Image 13

Fig. 10-5

5.Press button “M1” or “M2” to record the measurement in “M1” or “M2” memory, respectively. For example, if button “M2” is pressed, the display will alternate as shown in Fig. 10-5. If neither button “M1” nor “M2” is pressed, the measurement will not be recorded.

6.Press button “O/” to return to standby mode. Wait for at least 5 minutes before taking another measurement. If the device remains unused for 3 minutes, it will return to standby mode automatically.

Automatic Inflation

This device has 4 levels of inflation pressure: 190 mmHg, 230 mmHg, 270 mmHg and 300 mmHg. When 190 mmHg is insufficient to obtain accurate measurement, or arm movement occurs, the device

will automatically inflate to an appropriate pressure level in order to obtain the blood pressure measurement.

Rapid Deflation During Measurement

If you do not feel well during measurement, or want to stop measurement for any reason, press button “O/” to quickly deflate the cuff and return to standby mode.

13

Page 13
Image 13
Lumiscope 1133 instruction manual Automatic Inflation, Rapid Deflation During Measurement