15 TROUBLESHOOTING

SYMPTOM

CHECK POINT

REMEDY

 

 

 

No display when button “O/I”

The batteries need to be replaced

Replace the batteries with new ones

is pressed

Battery polarities are incorrectly positioned

Install the batteries in correct position

 

 

The battery contacts are corroded

Clean the battery terminals with a dry cloth

 

 

 

Inflation stops then starts

See <TAKING A MEASUREMENT/Automatic Inflation>

See <TAKING A MEASUREMENT/Automatic Inflation>

again

Did you talk or move your arm (or hand) during

Keep quiet and still during the measurement

 

 

measurement?

 

 

 

 

The reading is extremely low

Ensure the cuff is positioned at the same level

Ensure your posture is correct

or high

as the heart

 

 

Is the cuff wrapped correctly?

Ensure the cuff is wrapped correctly

 

Did you strain your arm (or hand) during

Relax during measurement

 

measurement?

 

 

Did you talk or move your arm (or hand) during

Keep quiet and still during the measurement

 

measurement?

 

 

 

 

Pulse rate is too low or too

Did you talk or move your arm (or hand) during

Keep quiet and still during the measurement

high

measurement?

 

 

Did you take your measurement right after

Take measurement again after resting for at least 5

 

exercise?

minutes

 

 

 

The batteries wear out too

Faulty batteries used

Suggestion: Use alkaline batteries of known

quickly

 

manufacturers

 

 

 

19