13  | 
  | 
  | 
  | ↑ and ↓ Keys- The ↑ and ↓ arrow keys are used in  | 
  | 
  | 
  | 
  | conjunction with the CALIBRATE key to calibrate the unit or in  | 
  | conjunction with the LOW SET and HIGH SET keys to adjust the alarm  | |||
  | settings. When used after pressing the CALIBRATE key, pressing the ↑  | |||
  | or ↓ key will raise or lower the displayed oxygen value in .1%  | |||
  | increments. When either of these keys are held down for more than 1  | |||
  | second, the display will scroll at a rate of .4% per second. When used  | |||
  | after pressing either the LOW SET or HIGH SET keys, pressing the ↑ or  | |||
  | ↓ key will raise or lower the displayed alarm setting in 1% increments. In  | |||
  | the alarm set mode, holding these keys down for more than 1 second  | |||
  | will cause the display to scroll at a rate of 4% per second.  | |||
14  | 
  | 
  | 
  | LOW SET Key- When the keypad is unlocked, pressing the  | 
  | 
  | 
  | ||
  | 
  | |||
  | 
  | 
  | ||
  | 
  | 
  | 
  | LOW SET key will cause the "LOW" icon to flash. The low  | 
  | 
  | 
  | 
  | |
  | alarm setting can then be changed using the ↑ and ↓ arrow keys. When  | |||
  | the low alarm value is set and the LOW SET or LOCK/UNLOCK key is  | |||
  | pressed again, the unit exits the alarm set mode and returns to normal  | |||
  | operation. The unit will also revert to normal operation (storing the most  | |||
  | recent alarm value) if 10 seconds elapse and no keys are pressed. The  | |||
  | ||||
  | alarm setting is adjustable to 15%.  | |||
15  | 
  | 
  | 
  | HIGH SET Key- When the keypad is unlocked, pressing the  | 
  | 
  | 
  | 
  | HIGH SET key will cause the "HIGH" icon to flash. The high  | 
  | alarm setting can then be changed using the ↑ and ↓ arrow keys. When  | |||
  | the high alarm value is set and the HIGH SET or LOCK/UNLOCK key is  | |||
  | pressed again, the unit exits the alarm set mode and returns to normal  | |||
  | operation. The unit will also revert to normal operation (storing the most  | |||
  | recent alarm value) if 10 seconds elapse and no keys are pressed.  | |||
16  | 
  | 
  | SILENT Key- In an alarm condition, pressing the SILENT key will  | |
  | 
  | 
  | deactivate the alarm beeper, but the alarm LED and display will  | |
  | continue to flash. In 120 seconds, the alarm beeper will sound again.  | |||
  | This process will be repeated until the alarm condition is cancelled. For  | |||
  | an alarm condition to be cancelled, the concentration must be .1% higher  | |||
  | than the alarm setting (in the presence of a low alarm condition) or .1%  | |||
  | lower than the alarm setting (in the presence of a high alarm condition).  | |||
D External Probe | ||||
17  | Sensor with Diverter- The sensor (with diverter) is designed to fit  | |||
industry standard, 15 mm I.D. "T" adapters.  | ||||
18  | Extendible Cable- The extendible cable allows the sensor to be  | |||
positioned up to 10 feet from the side of the unit.  | ||||
5