ERROR MESSAGES
The blood pressure monitor registers any problems and shows the following error messages on the display:
Error message | Meaning | Cause | |
Error during | 1. | The armband is not correctly attached. | |
| pumping | 2. | The armband is broken. |
|
| 3. | The pump or valves are defective. |
Disturbances or | 1. | The user moves or speaks | |
| noise during readings | during the reading. | |
|
| 2. | The user is too nervous. |
Error while emptying | 1. | The valves are defective. | |
| the armband | 2. | The armband is not correctly attached. |
Abnormal reading | The user may have a heart problem. | ||
HI | The pressure in the | The air tube is blocked, the pump | |
| armband is too high. | mechanism is defective or the pressure | |
|
| monitor is defective. | |
Battery symbol | Flat batteries | The batteries are almost dead. |
Solution
1.Attach the armband correctly.
2.Contact the retailer.
3.Contact the retailer.
1.The user must sit still and avoid talking during the reading.
2.Have a blood pressure reading taken by a doctor.
1.Contact the retailer.
2.Attach the armband correctly.
Consult a doctor.
Try to take another reading, or contact the retailer.
Replace the batteries.
35