![](/images/new-backgrounds/1058223/5822381x1.webp)
Programming special functions
3. Setting the buzzer
A buzzer can be set to signal the following:
–End of program = continuous buzzer
–Error = rhythmic buzz
–Timer to allow sampling of the last final rinse water
= intervals of 3 short buzzes and
2 seconds break
^Turn the program selector to "STOP" f.
^Turn the machine off.
^Press and hold "Display" ß and "Start" 6. At the same time, turn the lab washer on with the "On/Off" button.
The displays shows the current program status "P...".
The "Fill/Drain" indicator lights.
^Press "Drying" 0 4 times. "E04" appears in the display.
^Turn the program selector to "Inorganica". A number between "20" and "27" appears in the display.
"20" = Buzzer is turned off "21" = End of program
(factory setting) "22" = Error
"23" = Error and End of program "24" = Test sample extraction "25" = End of program and
Test sample extraction "26" = Error and Test sample
extraction
"27" = All functions
(End of program, Error, Test sample extraction).
^Press and hold "Display" ß until the desired value appears in the display.
^Press the "Start" 6 button. "SP" appears in the display.
^Press the "Start" 6 button again. The change has been saved.
^Please note the changed setting:
To turn the buzzer off:
^Turn the program selector to "STOP" f or press the "Display" ß button.
41