Adjust the SLOPE trimmer (pH4/10) on the front panel, with a small screwdriver, until the LCD shows the

pH value of the buffer at the temperature of measurement (see the pH versus temperature chart).

E.g. in this case, if the temperature is 15ºC, the meter display should be

adjusted to read “pH 4.00” (or pH 10.01 would be adjusted to 10.12 pH).

Calibration is now complete.

pH VERSUS TEMPERATURE CHART:

TEMP

 

pH VALUES

 

 

 

 

 

 

 

 

°C

°F

MA9004

 

MA9007

 

MA9010

 

 

 

 

 

 

 

0

32

4.01

 

7.13

 

10.32

 

 

 

 

 

 

 

5

41

4.00

 

7.10

 

10.24

 

 

 

 

 

 

 

10

50

4.00

 

7.07

 

10.18

 

 

 

 

 

 

 

15

59

4.00

 

7.04

 

10.12

 

 

 

 

 

 

 

20

68

4.00

 

7.03

 

10.06

 

 

 

 

 

 

 

25

77

4.01

 

7.01

 

10.01

 

 

 

 

 

 

 

30

86

4.02

 

7.00

 

9.96

 

 

 

 

 

 

 

35

95

4.03

 

6.99

 

9.92

 

 

 

 

 

 

 

40

104

4.04

 

6.98

 

9.85

 

 

 

 

 

 

 

45

113

4.05

 

6.98

 

9.85

 

 

 

 

 

 

 

50

122

4.06

 

6.98

 

9.82

 

 

 

 

 

 

 

55

131

4.07

 

6.98

 

9.79

 

 

 

 

 

 

 

60

140

4.09

 

6.98

 

9.77

 

 

 

 

 

 

 

65

149

4.11

 

6.99

 

9.76

 

 

 

 

 

 

 

70

158

4.12

 

6.99

 

9.75

 

 

 

 

 

 

 

Milwaukee

BATTERY REPLACEMENT:

When the battery becomes weak the meter will display “”.

When the low battery indicator appears, only a few hours of

battery

life remain.

A low

battery will

result

in unreliable

m e a s u r e m e n t s .

P r o m p t b a t t e r y

r e p l a c e m e n t i s

required .

Battery replacement must only take place

in a non-hazardous area using an alkaline 9V battery.

Turn the meter off, slide the battery compartment cover located at the rear of the meter off and replace the 9V battery with a new one. Make sure the battery contacts are fully engaged in the connector, seat the battery in its compartment and replace the cover.

OPTIONAL ACCESSORIES:

 

 

MA9004

pH4.01 buffer

solution,

 

220

mL bottle

 

 

MA9007

pH7.01 buffer

solution,

 

220

mL bottle

 

 

MA9010

pH10.01

buffer

solution,

 

220

mL bottle

 

 

MA9015

Electrode storage solution,

 

220

mL bottle

 

 

MA9016

General cleaning solution,

 

220

mL bottle

 

 

M10000B

Rinse solution, 20 mL

 

sachet (25 pcs.)

 

MA911B/1

pH

electrode with

BNC

 

connector and 1 m cable

MA950

Portable

meter

wall

 

mounting

kit

 

 

SPECIFICATIONS:

 

 

MW 101

RANGE

0.00 to 14.00 pH

RESOLUTION

0.01 pH

ACCURACY (@25°C)

±0.02 pH

TEMPERTURE

Manual setting

COMPENSATION

0 to 50ºC

CALIBRATION

Manual, 2-point

with use of Offset and Slope trimmers

pH ELECTRODE

MA911B/1 (included)

ENVIRONMENT 0 to 50°C, 95% RH max. BATTERY TYPE 1 x 9V alkaline (included)

BATTERY LIFE

approx. 70 hours of use

DIMENSIONS

 

143 x 80 x 32 mm

WEIGHT

220 g (with battery) meter only

ISTMW101 01/10