Note: Pressing the CAL key again will change the required buffer value, making possible a single point cali- bration at pH 4.01 or pH 10.01.

Wait till the hourglass symbol on the display turns off and the “pH” tag stops

blinking, then press the CFM key to confirm the off- set calibration. “pH 4.01” will flash on the LCD.

Rinse the pH electrode and the temperature

probe in the first pH4.01 beaker, then immerse them into the second pH4.01 beaker.

Follow the same procedure if using the

 

pH10.01 solution; select the correct

 

buffer by pressing the CAL key.

Wait till the hourglass symbol on the

 

LCD turns off, then press the CFM key

BATTERY REPLACEMENT:

When the battery becomes so weak that no reliable measurement can be guaran- teed, the meter will display “Eb” for a while and then turns off.

It is recommended to replace immediately the battery.

Battery replacement must only take place in a non-hazardous area using a 9V alka- line battery.

Turn the meter off, slide off the battery compart- ment cover at the rear of the meter and replace the 9V battery with a new one.

Make sure the battery contacts are tight and secure, seat the battery in its compartment and insert back the cover.

SPECIFICATIONS:

RANGE

-2.00 to 16.00 pH

 

-5 to 70°C

RESOLUTION

0.01 pH / 0.1°C

ACCURACY (@25°C)

±0.02 pH / ±0.5°C

TYPICAL EMC DEV.

±0.02 pH / ±0.5°C

TEMPERATURE

Automatic,

COMPENSATION

0 to 70°C

CALIBRATION Automatic, at 1 or 2 points

pH ELECTRODE

MA911B/1 (included)

TEMPERATURE

MA830R (included)

PROBE

 

ENVIRONMENT 0 to 50°C; RH 95% max. BATTERY TYPE 1 x 9V alkaline (included)

BATTERY LIFE

approx. 750 hours of use

AUTO-OFF

after 8 minutes of non-use

DIMENSIONS

145 x 80 x 40 mm

WEIGHT

220 g (with battery)

to confirm the slope calibration.

Notes:

To exit the calibration mode, turn the meter off by pressing the ON/OFF key.

To perform a single point calibration (offset) at pH7.01, press the ON/OFF key when the second buffer is asked.

After battery replacement, the meter dis- plays the “Clr” message and recalibration is needed.

M ilw a u ke e

OPTIONAL ACCESSORIES:

MA9004 pH4.01 buffer solution, 220 mL bottle

MA9007 pH7.01 buffer solution, 220 mL bottle

MA9010 pH10.01 buffer solution, 220 mL bottle

MA9015 Electrode storage solu- tion, 220 mL bottle

MA9016 General cleaning solu- tion, 220 mL bottle

M10000B Rinse solution, 20 ml sachet (25 pcs.)

MA911B/1 pH electrode with BNC connector and 1 m cable

MA830R Temperature probe

MA950Portable meter wall fix-

ing kit

ISMIL102R1 01/2003