MAXIMUM VALUE

Main Measurement Display Screen (Shaped Probes)

BAR GRAPH

FIELD STRENGTH

 

• Indicates approximate

• Indicates actual measured value

 

field strength

• Unit of measure is always % STD

• 100 segment, three

for shaped frequency response

probes

10:1 (10 dB) ranges

 

Indicates the maximum value that has occurred since you began making measurements or since resetting the maximum using

F2 (MAXR)

The unit of measure is the same as for the instantaneous value shown on the line above.

12 . 1 % S T D

27 . 5 M A X C F 1 . 0 0

A L T M A X R M E N U F R E Q

F1 F2 F3 F4

F1

 

 

F3

 

 

 

 

 

 

 

 

 

 

Change to the

Go to the First

Alternate

Menu Screen

Measurement

 

Display Screen

 

 

F4

F2

Allows you to enter

 

Reset the

or change the

Maximum Value to

frequency that you

the current Field

are correcting for

Strength displayed

probe frequency

on the second line

deviation

PROBE FREQUENCY CORRECTION

CF = 1.00 indicates no correction factor is being applied

CF= X.XX indicates the numeric value that is being applied

A frequency displayed here indicates that the meter is applying a correction factor based on the stored calibration data for the specific probe in use.

Main Measurement Display Screen (Flat Probes)

The Main Measurement Display Screen for flat frequency response probes is the same as the screen for shaped frequency response probes except as indicated below.

FIELD STRENGTH

Indicates actual measured value

Unit of measure can be changed by going to the First Menu Screen

6. 2 6 m W / c m 2

11 . 5 M A X C F 1 . 0 0

A L T M A X R M E N U F R E Q

F1 F2 F3 F4

Chapter 4 Making a Measurement

19