Alternate Measurement Display Screen (Shaped Probes)
BAR GRAPH | FIELD STRENGTH | |
| ||
• Indicates approximate | • Indicates actual measured value | |
| ||
field strength | • Unit of measure is always % STD | |
• 100 segment, three 10:1 | for shaped frequency response | |
probes | ||
(10 dB) ranges | ||
|
CALCULATED FIELD STRENGTH
•Displays field strength in mW/cm2 that has been calculated from the % STD reading using the referenced standard and the indicated frequency
12 . 1 % S T D
0. 6 1 m W / c m 2 1 0 . 0 0 G H z
M A I N M A X R M E N U | F R E Q | ||||||
| F1 | F2 |
| F3 | F4 | ||
F1 |
|
| F3 |
|
|
| |
|
|
|
|
|
| ||
|
|
|
|
|
| ||
Change to the |
| Go to the First |
| ||||
Alternate |
| Menu Screen |
|
Measurement
Display Screen
F4
Allows you to enter or change the frequency that you are correcting for probe frequency deviation
PROBE FREQUENCY CORRECTION
•CF = 1.00 indicates no correction factor is being applied
•CF= X.XX indicates the numeric value that is being applied
•A frequency displayed here indicates that the meter is applying a correction factor based on the stored calibration data for the specific probe in use.
CALCULATED PERCENT OF STANDARD
Alternate Measurement Display Screen (Flat Probes)
The Alternate Measurement Display Screen for flat frequency response probes is the same as the screen for shaped frequency response probes except as indicated below.
•Displays percent of standard that has been calculated from the field strength reading using the referenced standard and the indicated frequency
6. 2 6 m W / c m 2
12 5 . 2 % S T D 1 0 . 0 0 G H z
M A I N M A X R M E N U F R E Q
F1 F2 F3 F4
FIELD STRENGTH
•Indicates actual measured value
•Unit of measure can be changed by going to the First Menu Screen