To set the meter for a specific frequency...

1.Press F4 (FREQ) from the measurement screen. A new screen will appear to allow you to enter a frequency.

F R E Q U E N C Y : _ _ . _ _ G H

C O R R E C T . F A C T O R : 1 . 0 0

k H z M H z G H z S A V E

F1 F2 F3 F4

2.Enter the frequency that you want to use for the time- averaged measurements. Make sure you fill in all four positions. If you make a mistake just enter more digits until correct. For greater accuracy, use the units that allow the larger number. For example, enter

10.00 kHz, not 00.01 MHz.

3.Change the units (if needed) by pressing either F1 (kHz), F2 (MHz), or F3 (GHz).

4.Press F4 (SAVE). The meter will automatically return to the measurement screen.

5.If the probe used is a probe in memory, the frequency that you have set will be displayed on the right side of the third line of the display. The meter will also automatically correct the readings for probe frequency deviation. The meter does this by using the closest calibration frequencies above and below the selected frequency to calculate and apply an appropriate correction factor. This new correction factor is briefly shown on the second line of the display after you press F4 (SAVE) before the meter automatically returns to the measurement screen.

62

Operating the Meter