Note
If you access the menu system from the Opening Menu Screen, you will return to that screen. Once a probe has been selected and zeroed, the measurement screen will reflect the averaging mode that you have selected.
7.The meter will continue to make real time measurements until you initiate making the time- averaged measurement.
8.Press F1 (START) when you are ready to begin the
9.The averaged value is displayed on the screen and will flash every second each time it is updated. The meter will automatically stop at the end of the averaging interval, which was determined by the standard and the frequency.
10.To end time averaging prior to the end of the averaging interval, press F4 (STOP).
8. 7 1 m W / c m 2 A V E R A G E
11 . M A X 1 0 . 0 0 G H z
|
| P A U S E | S T O P |
F1 | F2 | F3 | F4 |
11.At the end of the averaging interval, the meter will show the
8 . 7 1 m W / c m 2 | A V E R A G E |
11 . 5 M A X 1 0 . 0 0 G H z
S T A R T C L R M E N U Q U I T
F1 | F2 | F3 | F4 |
12.To make another
64 | Operating the Meter |