© National Instruments Corporation 33 NI 5441 Specifications
Specification Valu e Comments
Digital Performance
Maximum
NCO Spur
< –90 dBc Full-Scale Output
FIR
Interpolation
IQ Rate Range
(with 100 MS/s
Sample Clock
Rate)
OSP Out of Band
Suppression
OSP Passband
Ripple
2195 kS/s to
8.33 MS/s
63 dB 0 to –0.08 dB FIR Filter Type =
Flat. Passband =
0.4. Ripple
Measurement to
0.4 × IQ Rate.
Stop Band
Suppression from
0.6 × IQ Rate.
497.6 kS/s to
4.16 MS/s
74 dB 0 to –0.08 dB
848.8 kS/s to
2.08 MS/s
40 dB 0 to –0.8 dB