Chapter 3 Hardware Overview
DAQCard E Series User Manual 3-10
National Instruments Corporation
In below-low-level analog triggering mode, the trigger is generated
when the signal value is less than
lowValue
.
HighValue
is unused.
In above-high-level analog triggering mode, the trigger is
generated when the signal value is greater than
highValue
.
LowValue
is unused.
Figure 3-5.
Below-Low-Level Analog Triggering Mode
Figure 3-6.
Above-High-Level Analog Triggering Mode
lowValue
Trigger
highValue
Trigger