NI 6509 User Guide and Specifications 22 ni.com
Shock and Vibration (PXI-6509 Only)Operational shock...................................30 g peak, half-sine, 11 ms pulse
(tested in accordance with
IEC-60068-2-27; test profile
developed in accordance with
MIL-PRF-28800F)
Random vibration
5 to 500 Hz, 0.3 grms.....................Operating
5 to 500 Hz, 2.4 grms.....................Nonoperating
Random vibration is tested in accordance with IEC-60068-2-64. The
nonoperating test profile exceeds the requirements of MIL-PRF-28800F,
Class 3. Random vibration is tested in accordance with IEC-60068-2-64.
The nonoperating test profile exceeds the requirements of
MIL-PRF-28800F, Class 3.
SafetyThis product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or visit ni.com/
certification, search by model number or product line, and click the appropriate link
in the Certification column.
Electromagnetic CompatibilityThis product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement, control,
and laboratory use:
• EN 61326 EMC requirements; Minimum Immunity
• EN 55011 Emissions; Group 1, Class A
• CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A
Note For EMC compliance, operate this device according to product documentation.