Data rate range (fs) using external master timebase

Minimum....................................

390.6 S/s

Maximum ...................................

51.2 kS/s

Data rates1 (f )...................................

fM ⎟ 256

-------------------- , n = 1, 2, ..., 31

s

n

 

Operating voltage ranges

 

Minimum....................................

294 Vrms

Typical ........................................

300 Vrms

Typical scaling coefficient ................

50.66 μV/LSB

Overvoltage protection .....................

±450 VDC

Input coupling...................................

DC

Input impedance (AI+ to AI–) ..........

1 MΩ

1The data rate must remain within the appropriate data rate range. Refer to the Understanding NI 9225E Data Rates section for more information.

NI 9225E Operating Instructions and Specifications 14

ni.com