© National Instruments Corp. 7 NI 9233 Operating Instructions and Specifications
Each channel has a BNC connector to which you can connect an
Integrated Electronic Piezoelectric (IEPE) sensor. The center pin
of the connector, AI+, provides the DC excitation and the AC
signal connection. The shell of the connector, AI–, provides the
excitation return path and the AC signal ground reference.
You can connect ground-referenced or floating IEPE sensors to the
NI 9233. You can avoid picking up ground noise by using a floating
connection. Typical IEPE sensors have a case that is electrically
isolated from the IEPE electronics, so connecting the sensor to the
NI 9233 results in a floating connection even though the case of the
sensor is grounded. To further minimize ground noise, prevent
themetal shells of the BNC connectors from touching each other,
the modules, or the chassis.
If you make a ground-referenced connection between the IEPE
sensor and the NI 9233, make sure the voltage on the AI– shell
isin the common-mode range to ensure proper operation of the
NI9233. The AI– shell is protected against accidental contact
withovervoltages within the overvoltage protection range. Refer
tothe Specifications section for more information about operating
voltages and overvoltage protection. Figures2 and 3 illustrate
connecting grounded and floating IEPE sensors to the NI 9233.