© National Instruments Corp. 19 NI 9852 Operating Instructions
Isolation VoltagesPort-to-port
Withstand....................................500 Vrms, 5s
Continuous.......... .......................60 VDC,
Measurement Category I
Port-to-earth ground
Withstand....................................500 Vrms, 5s
Continuous.......... .......................60 VDC,
Measurement Category I
Safety StandardsThe NI 9852 is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement,
control, and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the
product label, or visit ni.com/certification, search
by model number or product line, and click the
appropriate link in the Certification column.