Chapter 2 Connecting Signals
NI PXI-4224 User Manual 2-12 ni.com

Ground-Referenced Signal Connection

Figures 2-3, 2-5, 2-7, and 2-9 illustrate the ground-referenced signal
connection. In this configuration, the voltage source being measured is
referenced to its own ground reference that is connected through a
conductive path to the instrument ground reference. For example, the path
can be through a common earth ground or through the power line ground.
To connect a ground-reference signal source to the NI PXI-4224, the signal
(VSIG +) is connected to the NI PXI-4224 channel (AI X+). The signal
reference (VSIG –) is connected to the channel reference (AI X–).

Shielded Ground-Referenced Signal Connection

(Recommended)

Figures 2-5 and 2-9 illustrate shielded ground-referenced signal
connections. The connection to this signal source is identical to the
ground-referenced signal connection with the addition of shielding
around the field wiring. The shielding is grounded at the signal source
ground (VSIG Ground Reference). Connect the signal (VSIG +) to the
NI PXI-4224 channel (AI X+). Connect the signal reference (VSIG –) to the
channel reference (AI X–).
This shielding scheme is effective at reducing capacitive or electrically
coupled noise. The same concerns regarding the difference in ground
potentials, discussed in the Ground-Referenced Signal Connection section,
also apply to this configuration.
For more information about the function of the NI PXI-4224 and other
measurement considerations, refer to Chapter 4, Theory of Operation.