Appendix A Specifications for
© National Instruments Corporation A-5 NI PXI-4224 User Manual
Weight.................................................... 279 g (9.8 oz)
Analog input signal connector ............... 25-Pin D-SUB

Maximum Working Voltage

(Signal + common-mode) each input should remain within 42.4 Vpeak or
60 VDC of ground.
Maximum working voltage refers to the signal voltage plus the CMV.
Channel-to-earth (inputs) ....................... 42.4 Vpeak or 60 VDC,
Measurement Category I
Channel-to-channel (inputs)................... 42.4 Vpeak or 60 VDC,
Measurement Category I
Caution This device is rated for Measurement Category I and is intended to carry signal
voltages no greater than 42.4Vpeak or 60 VDC. Do not use this device for connection to
signals or for measurements within Categories II, III, or IV.

Isolation Voltages

Channel-to-channel, channel-to-earth isolation
Continuous ...................................... 60 VDC,
Measurement Category I
Withstand ........................................ 850 Vrms verified by a 5 s
dielectric withstand type test
Channel-to-bus
Continuous ...................................... 60 VDC,
Measurement Category I
Withstand ........................................ 1400 Vrms verified by a 5 s
dielectric withstand type test

Environmental

Operating temperature............................ 0 to 55 °C
Storage temperature ............................... 40 to 70 °C
Humidity ................................................ 10 to 90% RH, noncondensing
Maximum altitude .................................. 2,000 m
Pollution Degree (indoor use only) ........ 2