NI USB-9219 User Guide and Specifications 10 ni.com
Wiring the NI USB-9219 Device
Caution A high voltage screw terminal backshell must be installed when using hazardous
voltages (>42.4 Vpk, 60 VDC).
Note Table2 illustrates the accessories available from ni.com for use with the
NIUSB-921 9.
The NI USB-9219 has four 6-terminal connectors that provide connections
for four analog input channels. Connect the positive signal of the signal
source to the positive input signal terminal (HI) and the negative signal of
the signal source to the negative input signal terminal (LO). Use the
excitation terminals if your sensor requires a separate excitation
connection. Refer to Table3 for the signal names and Table 4 for the
terminal assignments for each mode. Refer to the NI USB-9219 Circuitry
section for information about connections in each mode.
Table 2. Accessories
Accessory Accessory Description
NI 9972 Strain relief for 6-position spring terminal
NI 9973 6-position connectors
Table 3. Signal Names
Module Terminal Signal Name Signal Description
1T+ TEDS Data
2T– TEDS COM
3EX+/HI*Positive excitation or input
signal
4HI Positive input signal
5EX–/LO*Negative excitation or input
signal
6LO Negative input signal
* Depending on the mode, terminals 3 and 5 are either the excitation signals or the input signals.
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
Ch 0
Ch 1
Ch 2
Ch 3