NI USB-9219 User Guide and Specifications 22 ni.com
SafetyIf you need to clean the module, wipe it with a dry towel.
Safety VoltagesConnect only voltages that are within these limits.
Isolation
Channel-to-channel
Continuous...............................250 VAC,
Measurement Category II
Withstand.................................1390 VAC, verified by a
5 s dielectric withstand test
Channel-to-earth ground
Continuous...............................250 VAC,
Measurement Category II
Withstand.................................2300 VAC, verified by a
5 s dielectric withstand test
Measurement Category II is for measurements performed on circuits
directly connected to the electrical distribution system. This category refers
to local-level electrical distribution, such as that provided by a standard
wall outlet, for example, 115 V for U.S. or 230 V for Europe. Do not
connect the NI 9219 to signals or use for measurements within
Measurement Categories III or IV.
Safety StandardsThis product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or visit ni.com/
certification, search by module number or product line, and click the appropriate link
in the Certification column.
Hazardous LocationsThe USB-9219 is not certified for use in hazardous locations.