© National Instruments Corporation 5 SCC-TC Series Thermocouple Input Modules User Guide
The SCC-TC02 has a fixed screw-terminal receptacle and a removable
screw-terminal block, as shown in Figure2.
Figure 2. SCC-TC02 Two-Part Screw-Terminal System
After you install the SCC-TC02, attach the input signals to the screw
terminals of the module.
1. Remove power from the signal lines.
2. Strip 7 mm (0.28 in.) of insulation from the ends of the signal wires.
3. Insert the wires into the screw terminals.
4. Tighten the screws to 0.5 to 0.6 Nm (4.4 to 5.3 lb in.) of torque.
The SCC-TC02 accepts up to three signals: TC+, TC–, and GND. TC+ is
the positive thermocouple lead and TC– is the negative thermocouple lead.1
The GND terminal connects to AIGND on the E Series DAQ device.
The SCC-TC0X has a 10 M bias resistor connected from the negative
thermocouple input to ground. This resistor allows the thermocouple to be
ground-referenced or floating without requiring external bias resistors
connected to ground.
1 SCC Screw-Terminal Receptacle 2 Removable Screw-Terminal Block
1 For ANSI color-coded J-type thermocouples, the red wire is negative and the white wire is positive. Refer to the thermocouple
data sheet if possible. You can find information about other color-coding schemes in the NI KnowledgeBase at ni.com/
support.