2.2Electrical

OMEGA CDE-45PConductivity Sensor

Part 2– Installation

2.2Electrical

The Model CDE-45P sensor comes standard with 15 feet of 6 conductor double shielded cable. The cable is permanently attached to the sensor, and a PEEK cordgrip is used to seal around the cable. Nevertheless, the cable should always be kept as clean and dry as possible.

DANGER: DO NOT connect sensor cable to power lines.

Serious injury may result.

Take care to route sensor cable away from AC power lines, adjustable frequency drives, motors, or other noisy electrical signal lines. Do not run signal lines in the same conduit as AC power lines. Run signal cable in dedicated metal conduit if possible. For optimum electrical noise protection, run an earth ground wire to the ground terminal in the transmitter

Refer to Figure 2-2, Cable Description and Figure 2-3, Wiring

Diagram for illustrative details on electrical installation.

Figure 2-2 Cable Description, Model CDE-45P

Only 6-wire shielded interconnect cable must be used when connecting the Model CDE-45P sensor to the analyzer. This high- performance, double shielded, polyethylene jacketed cable is specially designed to provide the proper signal shielding for the sensor used in this system. No substitutions can be made. Substituted cables may cause problems with system performance