OMEGA CDE-45PConductivity Sensor

Part 2– Installation

Part 2 – Installation

2.1General

The CDE-45P Conductivity Sensor is designed for industrial and municipal process applications. Mounting options include flow-through, submersion, insertion (special hardware required), or integral mount to the CDTX-45 Conductivity Monitor/Analyzer. The sensor-to-analyzer distance must not exceed 60 feet (18.2 meters).

Calibrate the sensor before placing it into the process.

See Model CDTX-45 Monitor/Analyzer Instruction

Manual for detailed calibration instructions.

The sensor comes with a removable guard that surrounds the electrode face. This guard minimizes interference effects in tight locations where the sensor face is close to surrounding objects. If it is removed, take care to leave at least 1 in3 of space in front of the electrodes. If the guard is to be used, the sensor must be calibrated with it in place, since the guard affects the sensor cell constant.

O&M Manual

- 5 -

Page 6
Image 6
Omega Engineering CDE-45P manual Part 2 Installation