0-5VDC: Measured Temp = [(Voltage Output ) x (T2 - T1)]+ T1
5
0-10VDC: Measured Temp = [(Voltage Output) x (T2 - T1)]+ T1
10
Where, T1 is the minimum temperature range
T2 is the maximum temperature range
3.5 - Resetting Temperature Values
The calculated temperature values (Min, Max, and Diff) can be reset at
any time by pressing the key in either the MAX, MIN, or DIF display
modes. This will reset the calculated temperatures as follows:
MAX = Current Temperature
MIN = Current Temperature
DIF = 0
After the reset, the unit starts to keep track of the Maximum, Minimum,
and Differential temperatures.
The following figures show the Main Housing, Optical Assemblies,
Mounting Bracket, and Air purge Collar.
Figure 4 - NEMA-4 Aluminum Enclosure
OS1592 Infrared Fiber Optic Thermometer/Transmitter