5. Accessories
Dimensions:

5. Accessories

Part no.

Description

Material

 

 

 

 

PH-31542-TC

Sensor

PP

 

 

cap

 

 

 

 

 

FPP-1220-0021

Sensor O-ring,

Viton®

FPP-1224-0021

2 required

EPR

FPP-1228-0021

 

Kalrez

 

 

 

 

CAUTION!

When replacing O-rings, apply O-ring lubricant to sensor O-rings prior to preamplifier/electrode assembly. Unlubricated O-rings may score the preamplifier sealing surface.

6. Specifications

PH-2720-PA pH/ORP Preamplifier

 

Housing material:

CPVC

 

Compatible sensors:

PHE-3271

Flat Surface pH Electrode

 

PHE-2716

Bulb pH Electrode

 

ORE-2715 Flat Surface ORP Electrode

 

ORE-2717 Bulb ORP Electrode

Input Impedance:

>1011

 

Operating temp.:

0 to 80 ° C

 

Gain:

X1 (unity)

 

Input power:

± 4.5 to ± 8 VDC, dual supply

Current consumption:

<1 mA, dual supply

Quality standard:

CE

 

3/4 in. NPT female threads

Std.15 ft/ 6. Specifications 4.6 m cable 7. Preamplifier TroubleshootingManual background

3.6 in./

91 mm

Manual backgroundManual background1.8 in./Manual backgroundManual background

46 mm

PHE Series pH Electrodes - ORE Series ORP Electrodes

General Specifications

 

Wetted parts:

 

Sensor body:

CPVC

O-rings:

Viton®

Electrode junction:

Porous UHMW polyethylene

Quality standard:

CE

Maximum pressure/temperature ratings:

7 bar (100 psi) max. @ 65 ° C (149 ° F)

4 bar (58 psi) max. @ 85 ° C (185 ° F)

BAR PSI

7100

575

350

225

°F 77 122 167 212

°C 25 50 75 100

SAFETY INSTRUCTIONS FOR IN-LINE ELEC- TRODE INSTALLATION

1. Do not remove from pressurized lines.

2. Do not exceed maximum temperature/pressure specifications.

3. Do not install/service without following installation instructions (see sensor manual).

4. Wear safety goggles and faceshield during installation/service.

5. Do not alter product construction.

6. Failure to follow safety instructions could result in severe personal injury!

7. Preamplifier Troubleshooting

Procedure:

A.Install sensor adapter into preamplifier

B.With preamplifier and instrument connected, simulate pH 4, 7, and 10 and record displayed mV readings (right) and approximate preamplifier response time between simulations.

C.Refer to table on next page for preamplifier troubleshooting tips.

Simulator

Actual displayed

Input

mV

4.00, +177 mV

________________________

7.00, 0 mV

________________________

10.00, -177 mV

________________________

Preamplfier response time:

________________________