4. Troubleshooting and Maintenance
Error Display | Cause | Remedy |
|
|
|
| Cuff is under inflated. | Carefully read and repeat the steps listed under |
| section 3.3. | |
|
| |
|
|
|
|
| Repeat measurement. Remain still and do not |
| Movement during measurement. | talk during measurement. |
|
| Refer to section 3.3. |
|
|
|
| Air plug disconnected. | Insert the air plug securely. |
| Refer to section 3.1. | |
|
| |
|
|
|
| Arm cuff not applied correctly. | Apply the arm cuff correctly. |
| Refer to section 3.1. | |
|
| |
|
|
|
| Clothing is interfering with the arm cuff. | Remove any clothing interfering with the arm cuff. |
| Refer to section 3.1. | |
|
| |
|
|
|
| Air is leaking from the arm cuff. | Replace cuff with new one. |
| Refer to Chapter 5. | |
|
| |
|
|
|
| The arm cuff was inflated above 299 | Do not inflate the arm cuff above 299 mmHg. |
| mmHg when inflating the cuff manually. | Refer to section 3.3. |
Device error. | Contact your OMRON retail outlet or distributor. |
EN |
Note:
The irregular heartbeat symbol ( ) may also be displayed with error messages.