4. Troubleshooting and Maintenance

Error Display

Cause

Remedy

 

 

 

 

Cuff is under inflated.

Carefully read and repeat the steps listed under

 

section 3.3.

 

 

 

 

 

 

 

Repeat measurement. Remain still and do not

 

Movement during measurement.

talk during measurement.

 

 

Refer to section 3.3.

 

 

 

 

Air plug disconnected.

Insert the air plug securely.

 

Refer to section 3.1.

 

 

 

 

 

 

Arm cuff not applied correctly.

Apply the arm cuff correctly.

 

Refer to section 3.1.

 

 

 

 

 

 

Clothing is interfering with the arm cuff.

Remove any clothing interfering with the arm cuff.

 

Refer to section 3.1.

 

 

 

 

 

 

Air is leaking from the arm cuff.

Replace cuff with new one.

 

Refer to Chapter 5.

 

 

 

 

 

 

The arm cuff was inflated above 299

Do not inflate the arm cuff above 299 mmHg.

 

mmHg when inflating the cuff manually.

Refer to section 3.3.

Device error.

Contact your OMRON retail outlet or distributor.