6.Specifications

Product description

Automatic Blood Pressure Monitor

Model

OMRON M6 AC (HEM-7322-E)

Display

LCD digital display

Measurement method

Oscillometric method

Measurement range

Pressure: 0 to 299 mmHg

 

Pulse: 40 to 180 beats / min.

Accuracy

Pressure: ±3 mmHg

 

Pulse: ±5% of display reading

Inflation

Fuzzy-logic controlled by electric pump

Deflation

Automatic pressure release valve

Memory

100 measurements with date and time for each user (1 and 2)

Rating

DC6V 4W

Power source

4 “AA” batteries 1.5V or optional AC adapter

 

(Adapter S-9515336-9, INPUT AC100-240V 50/60Hz 0.12A)

 

(Adapter UK-9983666-5, INPUT AC100-240V 50/60Hz 15VA)

Battery life

Approximately 1000 measurements (using new alkaline batteries)

Applied part

 

 

 

 

= Type BF

 

 

 

 

 

Protection against electric shock

Internally powered ME equipment (When using only the batteries)

Operating temperature/ humidity

 

 

 

 

= Class II ME equipment (AC adapter)

 

 

 

 

 

 

 

 

 

+10°C to +40°C / 30 to 85% RH

Storage temperature/ humidity/ air pressure

-20°C to +60°C / 10 to 95% RH / 700-1060hPa

IP classification

IP 20

Weight

Monitor: Approximately 380g without batteries

Outer dimensions

Arm cuff: Approximately 170g

Monitor: Approximately 124 (w) mm 90 (h) mm 161 (l) mm

 

Arm cuff: Approximately 145 mm 594 mm (air tube: 750 mm)

Cuff circumference

22 to 42 cm

Cuff / Tube material

Nylon, polyester, polyvinyl chloride

Package contents

Monitor, arm cuff, instruction manual, storage case, battery set, AC adapter,