4. Error Messages and Troubleshooting

4.2Troubleshooting

Problem

Cause

Solution

 

 

 

 

Arm cuff is applied too loosely.

Apply the arm cuff tighter.

 

Refer to section 3.1.

 

 

 

 

 

 

Movement or talking during

Remain still and do not talk during

The measurement result is extremely

measurement.

measurement.

high (or low).

Refer to section 3.3.

 

 

 

 

 

Clothing is interfering with the arm

Remove any clothing interfering

 

with the arm cuff.

 

cuff.

 

Refer to section 3.1.

 

 

 

 

 

 

The air connector is not securely

Make sure that the air tube is

 

connected securely.

 

connected into the air jack.

 

Refer to section 3.1.

Arm cuff pressure does not rise.

 

 

 

 

Replace the arm cuff with a new

 

 

 

Air is leaking from the arm cuff.

one.

 

 

Refer to section 5.3.

 

 

 

 

 

Apply the arm cuff correctly so that

Arm cuff deflates too soon.

The arm cuff is loose.

it is firmly wrapped around the arm.

 

 

Refer to section 3.1.

 

 

 

 

 

Inflate the arm cuff so that it is 30 to

Cannot measure or the results are

The arm cuff has not been inflated

40 mmHg above your previous

too low or too high.

sufficiently.

measurement result.

 

 

Refer to section 3.3.

 

 

 

EN

21

Page 26
Image 26
Omron M6 AC instruction manual Troubleshooting, Problem Cause Solution