How to use
Warning:
•Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before using them when as follows:
-the first time after purchase
-if the device has not been used for a long period of time
-if more than one person uses the same device
For the instructions, refer to “Cleaning and disinfecting” section.
1.Make sure that the power switch is in the off ( ) position.
2. Plug the power plug into a power outlet.
2
3. Remove the cap and the inhalation top from
the medication tank.
1
4. Add the correct amount of prescribed medication to the medication tank.
5. Make sure that the vaporiser head is securely attached to the inhalation top.
Cleaning and daily disinfecting
Cleaning
Clean the parts after each use to remove residual medication. This will avoid inefficient nebulization or infection.
■Washable parts
•Nebulizer Kit, Mask (PVC)/(SEBS), Mouthpiece, Nosepiece, Air Tube (Silicone)
Wash them in warm water and mild detergent (neutral detergent). Rinse them thoroughly with clean hot tap water and allow to air dry in a clean place.
■
• Compressor, Air Tube (PVC)
Firstly, make sure that the power plug is unplugged from the power outlet. Wipe clean with a soft cloth moistened with water or mild detergent (neutral detergent).
• Air Filter
Do not wash or clean the air filter. If the air filter becomes wet, replace it. Damp air filters can cause blockages.
Disinfecting
Always disinfect the parts after the last treatment of the day. If the parts are heavily stained, replace them with new ones.
To select a method for disinfection, refer to the table below.
o | : applicable X: not applicable |
Removing condensation from the air tube
If there is moisture or liquid remaining in the air tube, be sure to follow the procedure below to remove any moisture from within the air tube.
1)Make sure that the air tube is connected to the air connector on the compressor.
2)Remove the air tube from the nebulizer kit.
3)Turn on the compressor and pump air through the air tube to expel the moisture.
Changing the air filter
If the air filter has been used for more than 60 days, replace it with a new one.
1. Pull the air filter cover off the compressor.
2.Change the air filter.
Remove the old air filter with a sharp object and insert the new air filter.
3. Put the air filter cover back in place. Make sure it is attached properly.
Technical data
Name:
Model:
Type:
Rating:
Power Consumption:
Operating Temperature/ Humidity:
Storage and Transport
Temperature/Humidity/
Air Pressure:
Weight:
Dimensions:
Contents:
Classification:
= Class ll equipment
= Power off
COMP AIR
Compressor Nebulizer
+10°C to +40°C / 30% to 85%RH
Approx. 1.9kg (compressor only)
Approx. 170 (W) × 103 (H) ×182 (D) mm
Compressor, Nebulizer Kit, Air Tube (PVC, 200cm), Mouthpiece, Nosepiece, Adult Mask (PVC), Child Mask (PVC), 5 pcs Spare Air Filters, Carring Bag, Instruction Manual, Warranty Card.
Class ll equipment, Type B applied part
= Type B | Read the instruction |
applied part | manual carefully |
= Power on |
|
6. Put the inhalation top back inside the medication tank.
| Parts | Mouthpiece | Nosepiece | NebulizerKit | Air Tube (Silicone,100cm) (optional) | Adult(SEBS)Mask (optional) Child(SEBS)Mask (optional) | Adult (PVC)Mask Child (PVC)Mask | Air Tube (PVC,200cm) | Air Filter |
Disinfecting |
| ||||||||
|
|
|
|
|
|
|
|
| |
method |
|
|
|
|
|
|
|
|
|
Notes:
•Use only OMRON air filters designed for this device. Do not operate without a filter.
•Do not wash or clean the filter. If the air filter becomes wet, replace it. Damp air filters can cause blockages.
•There is no front or back orientation for the air filters.
•Check that air filters are clean and free of dust before inserting them.
Technical data for the OMRON
Particle Size: | **MMAD Approx. 3μm |
| MMAD = Mass Median Aerodynamic Diameter |
Medication Tank Capacity: | 7ml maximum |
Appropriate Medication | 2ml minimum - 7ml maximum |
Quantities: |
|
Sound: | *Noise level (at 1 m distance) 60dB |
7.Attach the cap and mask, mouthpiece or nosepiece to the nebulizer kit tightly.
Note:
It is not necessary to attach the cap when nebulizing with Adult mask. If the amount of aerosol is too great, attach the cap.
2
2
| 2 |
1 | 1 |
|
| Materials | PP | PP | PP (VaporiserHead:PC) | Tube:Silicone Connector:PP | Mask:SEBS Band:Rubber (nonlatex)- Adapter:PP | Mask:PVC (Phthalatefree) Band:Rubber (nonlatex)- | PVC (Phthalatefree) | Polyester |
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
| Alcohol | o | o | o | o | o | o | X | X |
| Disinfecting | ||||||||
| ethanol |
|
|
|
|
|
|
|
|
| Sodium |
|
|
|
|
|
|
|
|
| hypochlorite | o | o | o | o | o | o | X | X |
|
| ||||||||
| Milton* |
|
| ||||||
|
|
|
|
|
|
|
|
|
Troubleshooting
Check the following if the device should fail during operation.
Problem | Cause | Remedy |
|
|
|
Nothing happens when the | Is the power plug plugged into | Check that the plug is inserted |
power switch is pressed. | a power outlet? | in a power outlet. Unplug then |
|
| reinsert the plug if necessary. |
|
|
|
No nebulization or low | Is there medication in the | Add the correct amount of |
nebulization rate, when the | medication tank? | medication to the medication tank. |
power is on. |
|
|
| Is there too much/little |
|
| medication in the medication |
|
Nebulization Rate: | *Approx. 0.5ml/min (by weight loss) |
Aerosol Output: | **0.4ml (2ml, 1%NaF) |
Aerosol Output Rate: | **0.09ml/min (2ml, 1%NaF) |
Result of cascade impactor measurements for particle size**
Cumulative % particle mass of sodium fluoride undersize
| 100 | |
| 90 | |
Undersize% | 80 | |
70 | ||
60 | ||
50 | ||
|
8.Attach the air tube. Twist the air tube plug slightly and push it firmly into the air tube connector.
9.Hold the nebulizer kit as
indicated on the right.
Follow the instructions of your doctor or respiratory therapist.
Right angle
Caution:
Do not tilt the nebulizer kit at an angle of greater than 45 degrees in all directions. Medication may flow into the mouth or it may result in not nebulizing effectively.
10.Press the power switch to the on ( | ) position. As the compressor |
starts, nebulization begins and aerosol is generated. | |
Inhale the medication. |
|
11.When treatment is complete, turn the power off and unplug the compressor from the power outlet.
| Quaternary |
| o | o | o | o | o |
|
|
A | ammonium o | X | X | ||||||
| Osvan* |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Chlorhexidine | o | o | o | o | o | o | X | X |
|
| ||||||||
| Hibitane* |
|
| ||||||
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
| Amphoteric |
|
|
|
|
|
|
|
|
| Surfactant | o | o | o | o | o | o | X | X |
|
| ||||||||
| Tego* |
|
| ||||||
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
B | Boiling | o | o | o | o | o | X | X | X |
|
*an example of commercially available disinfectant.
A.Use a commercially available disinfectant. Follow the instructions provided by the disinfectant manufacturer.
Note: Never clean with benzene, thinner or a flammable chemical.
B.Parts may be boiled between 15 to 20 minutes.
After boiling, carefully remove the parts, shake
off excess water and allow to air dry in a clean environment.
Caution: Handling the Vaporiser Head
•Always wash the vaporiser head after each use.
•Do not use a brush or pin to clean the vaporiser head.
•When disinfecting the parts by boiling, be sure to boil them in plenty of water.
•Do not boil vaporiser head together with objects other than applicable nebulizer accessories.
| tank? |
|
|
|
|
| Is the vaporiser head missing | Attach the vaporiser head |
| or not assembled correctly? | correctly. |
|
|
|
| Is the nebulizer kit assembled | Assemble the nebulizer kit |
| correctly? | correctly. |
|
|
|
| Is the nozzle blocked? | Make sure that the nozzle is free |
|
| of blockages. |
|
|
|
| Is the nebulizer kit tilted at a | Make sure that the nebulizer kit is |
| sharp angle? | not tilted at an angle of more than |
|
| 45 degrees. |
|
|
|
| Is the air tube connected | Make sure that the air tube |
| correctly? | is correctly connected to the |
|
| compressor and nebulizer kit. |
|
|
|
| Is the air tube folded or | Make sure that the air tube does |
| damaged? | not contain kinks. |
|
|
|
| Is the air tube blocked? | Make sure that the air tube is free |
|
| of blockages. |
|
|
|
| Is the air filter dirty? | Replace the air filter with a new |
|
| one. |
|
|
|
The compressor is | Is the air filter cover attached | Attach the air filter cover correctly. |
abnormally loud. | correctly? |
|
|
|
|
The compressor is very hot. | Is the compressor covered? | Do not cover the compressor with |
|
| any type of cover during use. |
|
|
|
| Are the ventilation slots | Do not block the ventilation slots. |
| blocked? |
|
Note: If the suggested remedy does not solve the problem, do not try to repair the device
-no parts of the device are user serviceable. Return the device to an authorized OMRON retail outlet or distributor.
Cumulative |
|
| 40 |
| Individual tests | 30 | |
| Mean | 20 | |
|
| MMAD: 3μm | 10 |
|
|
| |
|
|
| 0 |
0.1 | 1 | 10 | 100 |
Particle Size Dp (μm)
*Measured by OMRON HEALTHCARE Co., Ltd.
**Independently measured by Prof. Dr. Hiroshi Takano, Department of Chemical
Engineering and Materials Science, Faculty of Science and Engineering, Doshisha University, Kyoto, Japan, according to EN
Notes:
•Subject to technical modification without prior notice.
•This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co., Ltd., Japan.
•The device may not work if the temperature and voltage conditions are different to those defined in the specifications.
•The device fulfils the provisions of the EC directive 93/42/EEC (Medical Device Directive) and the European Standard
•Performance may vary with drugs such as suspensions or high viscosity. See drug supplier’s data sheet for further details.
•See website of OMRON HEALTHCARE EUROPE to update technical information. URL: