How to use

Warning:

Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before using them when as follows:

-the first time after purchase

-if the device has not been used for a long period of time

-if more than one person uses the same device

For the instructions, refer to “Cleaning and disinfecting” section.

1.Make sure that the power switch is in the off ( ) position.

2. Plug the power plug into a power outlet.

2

3. Remove the cap and the inhalation top from

the medication tank.

1

4. Add the correct amount of prescribed medication to the medication tank.

5. Make sure that the vaporiser head is securely attached to the inhalation top.

Cleaning and daily disinfecting

Cleaning

Clean the parts after each use to remove residual medication. This will avoid inefficient nebulization or infection.

Washable parts

Nebulizer Kit, Mask (PVC)/(SEBS), Mouthpiece, Nosepiece, Air Tube (Silicone)

Wash them in warm water and mild detergent (neutral detergent). Rinse them thoroughly with clean hot tap water and allow to air dry in a clean place.

Non-washable parts

• Compressor, Air Tube (PVC)

Firstly, make sure that the power plug is unplugged from the power outlet. Wipe clean with a soft cloth moistened with water or mild detergent (neutral detergent).

• Air Filter

Do not wash or clean the air filter. If the air filter becomes wet, replace it. Damp air filters can cause blockages.

Disinfecting

Always disinfect the parts after the last treatment of the day. If the parts are heavily stained, replace them with new ones.

To select a method for disinfection, refer to the table below.

o

: applicable X: not applicable

Removing condensation from the air tube

If there is moisture or liquid remaining in the air tube, be sure to follow the procedure below to remove any moisture from within the air tube.

1)Make sure that the air tube is connected to the air connector on the compressor.

2)Remove the air tube from the nebulizer kit.

3)Turn on the compressor and pump air through the air tube to expel the moisture.

Changing the air filter

If the air filter has been used for more than 60 days, replace it with a new one.

1. Pull the air filter cover off the compressor.

2.Change the air filter.

Remove the old air filter with a sharp object and insert the new air filter.

3. Put the air filter cover back in place. Make sure it is attached properly.

Technical data

Name:

Model:

Type:

Rating:

Power Consumption:

Operating Temperature/ Humidity:

Storage and Transport

Temperature/Humidity/

Air Pressure:

Weight:

Dimensions:

Contents:

Classification:

= Class ll equipment

= Power off

COMP AIR

NE-C28P (NE-C28P-E)

Compressor Nebulizer

220-240V, 50/60Hz 192VA

+10°C to +40°C / 30% to 85%RH

-20°C to +60°C / 10% to 95%RH / 700 - 1060hPa

Approx. 1.9kg (compressor only)

Approx. 170 (W) × 103 (H) ×182 (D) mm

Compressor, Nebulizer Kit, Air Tube (PVC, 200cm), Mouthpiece, Nosepiece, Adult Mask (PVC), Child Mask (PVC), 5 pcs Spare Air Filters, Carring Bag, Instruction Manual, Warranty Card.

Class ll equipment, Type B applied part

= Type B

Read the instruction

applied part

manual carefully

= Power on

 

6. Put the inhalation top back inside the medication tank.

 

Parts

Mouthpiece

Nosepiece

NebulizerKit

Air Tube (Silicone,100cm) (optional)

Adult(SEBS)Mask (optional) Child(SEBS)Mask (optional)

Adult (PVC)Mask Child (PVC)Mask

Air Tube (PVC,200cm)

Air Filter

Disinfecting

 

 

 

 

 

 

 

 

 

 

method

 

 

 

 

 

 

 

 

 

Notes:

Use only OMRON air filters designed for this device. Do not operate without a filter.

Do not wash or clean the filter. If the air filter becomes wet, replace it. Damp air filters can cause blockages.

There is no front or back orientation for the air filters.

Check that air filters are clean and free of dust before inserting them.

Technical data for the OMRON NE-C28P Compressor with the V.V.T. Nebulizer Kit N-01:

Particle Size:

**MMAD Approx. 3μm

 

MMAD = Mass Median Aerodynamic Diameter

Medication Tank Capacity:

7ml maximum

Appropriate Medication

2ml minimum - 7ml maximum

Quantities:

 

Sound:

*Noise level (at 1 m distance) 60dB

7.Attach the cap and mask, mouthpiece or nosepiece to the nebulizer kit tightly.

Note:

It is not necessary to attach the cap when nebulizing with Adult mask. If the amount of aerosol is too great, attach the cap.

2

2

 

2

1

1

 

 

Materials

PP

PP

PP (VaporiserHead:PC)

Tube:Silicone Connector:PP

Mask:SEBS Band:Rubber (nonlatex)- Adapter:PP

Mask:PVC (Phthalatefree) Band:Rubber (nonlatex)-

PVC (Phthalatefree)

Polyester

 

 

 

 

 

 

 

 

 

 

 

 

 

Alcohol

o

o

o

o

o

o

X

X

 

Disinfecting

 

ethanol

 

 

 

 

 

 

 

 

 

Sodium

 

 

 

 

 

 

 

 

 

hypochlorite

o

o

o

o

o

o

X

X

 

 

 

Milton*

 

 

 

 

 

 

 

 

 

 

 

Troubleshooting

Check the following if the device should fail during operation.

Problem

Cause

Remedy

 

 

 

Nothing happens when the

Is the power plug plugged into

Check that the plug is inserted

power switch is pressed.

a power outlet?

in a power outlet. Unplug then

 

 

reinsert the plug if necessary.

 

 

 

No nebulization or low

Is there medication in the

Add the correct amount of

nebulization rate, when the

medication tank?

medication to the medication tank.

power is on.

 

 

 

Is there too much/little

 

 

medication in the medication

 

Nebulization Rate:

*Approx. 0.5ml/min (by weight loss)

Aerosol Output:

**0.4ml (2ml, 1%NaF)

Aerosol Output Rate:

**0.09ml/min (2ml, 1%NaF)

Result of cascade impactor measurements for particle size**

Cumulative % particle mass of sodium fluoride undersize

 

100

 

90

Undersize%

80

70

60

50

 

8.Attach the air tube. Twist the air tube plug slightly and push it firmly into the air tube connector.

9.Hold the nebulizer kit as

indicated on the right.

Follow the instructions of your doctor or respiratory therapist.

Right angle

Caution:

Do not tilt the nebulizer kit at an angle of greater than 45 degrees in all directions. Medication may flow into the mouth or it may result in not nebulizing effectively.

10.Press the power switch to the on (

) position. As the compressor

starts, nebulization begins and aerosol is generated.

Inhale the medication.

 

11.When treatment is complete, turn the power off and unplug the compressor from the power outlet.

 

Quaternary

 

o

o

o

o

o

 

 

A

ammonium o

X

X

 

Osvan*

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Chlorhexidine

o

o

o

o

o

o

X

X

 

 

 

Hibitane*

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Amphoteric

 

 

 

 

 

 

 

 

 

Surfactant

o

o

o

o

o

o

X

X

 

 

 

Tego*

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

B

Boiling

o

o

o

o

o

X

X

X

 

*an example of commercially available disinfectant.

A.Use a commercially available disinfectant. Follow the instructions provided by the disinfectant manufacturer.

Note: Never clean with benzene, thinner or a flammable chemical.

B.Parts may be boiled between 15 to 20 minutes.

After boiling, carefully remove the parts, shake

off excess water and allow to air dry in a clean environment.

Caution: Handling the Vaporiser Head

Always wash the vaporiser head after each use.

Do not use a brush or pin to clean the vaporiser head.

When disinfecting the parts by boiling, be sure to boil them in plenty of water.

Do not boil vaporiser head together with objects other than applicable nebulizer accessories.

 

tank?

 

 

 

 

 

Is the vaporiser head missing

Attach the vaporiser head

 

or not assembled correctly?

correctly.

 

 

 

 

Is the nebulizer kit assembled

Assemble the nebulizer kit

 

correctly?

correctly.

 

 

 

 

Is the nozzle blocked?

Make sure that the nozzle is free

 

 

of blockages.

 

 

 

 

Is the nebulizer kit tilted at a

Make sure that the nebulizer kit is

 

sharp angle?

not tilted at an angle of more than

 

 

45 degrees.

 

 

 

 

Is the air tube connected

Make sure that the air tube

 

correctly?

is correctly connected to the

 

 

compressor and nebulizer kit.

 

 

 

 

Is the air tube folded or

Make sure that the air tube does

 

damaged?

not contain kinks.

 

 

 

 

Is the air tube blocked?

Make sure that the air tube is free

 

 

of blockages.

 

 

 

 

Is the air filter dirty?

Replace the air filter with a new

 

 

one.

 

 

 

The compressor is

Is the air filter cover attached

Attach the air filter cover correctly.

abnormally loud.

correctly?

 

 

 

 

The compressor is very hot.

Is the compressor covered?

Do not cover the compressor with

 

 

any type of cover during use.

 

 

 

 

Are the ventilation slots

Do not block the ventilation slots.

 

blocked?

 

Note: If the suggested remedy does not solve the problem, do not try to repair the device

-no parts of the device are user serviceable. Return the device to an authorized OMRON retail outlet or distributor.

Cumulative

 

 

40

 

Individual tests

30

 

Mean

20

 

 

MMAD: 3μm

10

 

 

 

 

 

 

0

0.1

1

10

100

Particle Size Dp (μm)

*Measured by OMRON HEALTHCARE Co., Ltd.

**Independently measured by Prof. Dr. Hiroshi Takano, Department of Chemical

Engineering and Materials Science, Faculty of Science and Engineering, Doshisha University, Kyoto, Japan, according to EN 13544-1:2007+A1:2009.

Notes:

Subject to technical modification without prior notice.

This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co., Ltd., Japan.

The device may not work if the temperature and voltage conditions are different to those defined in the specifications.

The device fulfils the provisions of the EC directive 93/42/EEC (Medical Device Directive) and the European Standard EN13544-1:2007+A1:2009, Respiratory therapy equipment - Part1: Nebulizing systems and their components.

Performance may vary with drugs such as suspensions or high viscosity. See drug supplier’s data sheet for further details.

See website of OMRON HEALTHCARE EUROPE to update technical information. URL: www.omron-healthcare.com