Thank you for purchasing the OMRON COMP AIR.
This product was developed in conjunction with respiratory therapists for the successful treatment of asthma, chronic bronchitis, allergies and other respiratory disorders.
This is a medical device. Operate the device only as instructed by
Important safety instructions
Read all the information in the instruction manual and any other literature included in the box before using the device.
Warning: | Indicates a potentially hazardous situation which, if not avoided, could result in |
serious injury. |
(Usage)
General Safety Precautions:
• Inspect the device and parts before using them each time, and check that there are |
no problems. In particular, be sure to check the following: |
- That the nozzle or air tube are not damaged. |
- That the nozzle is not blocked. |
- That the compressor operates normally. |
• When using this device, there will be some noise and vibration caused by the pump in |
the compressor. There will also be some noise caused by the emission of compressed |
your doctor and/or respiratory therapist.
How the nebulizer kit works
The medication that is pumped up through the medication channel is mixed with compressed air which is generated by a compressor pump. The compressed air mixed with medication is turned into fine particles and sprayed when in contact with the baffle.
Baffle
AerosolAerosol
Medication
Nozzle Channel
Medication Medication
• For type, dose, and regime of medication follow the instructions of your doctor or |
respiratory therapist. |
• If you feel anything unusual during use, stop using the device immediately and |
consult your doctor. |
• Do not use only water in the nebulizer for inhaling purposes. |
• Keep the device out of the reach of unsupervised infants and children. The device |
may contain small pieces that can be swallowed. |
• Do not store the air tube while there is moisture or medication remaining inside. |
• Do not use or store the device where it may be exposed to noxious fumes or volatile |
substances. |
• Do not use the device where it may be exposed to flammable gas. |
• Do not cover the compressor with a blanket, towel, or any other type of cover during |
use. |
• Always dispose of any remaining medication after use, and use fresh medication each |
time. |
• Do not use in anaesthetic or ventilator breathing circuits. |
air from the nebulizer kit. This is normal and does not indicate a malfunction. |
• Operate the device only as intended. Do not use the device for any other purpose. |
• Do not use the device at temperatures greater than +40°C. |
• Make sure that the air tube is securely attached to the compressor and nebulizer kit, |
and does not come loose. Twist the air tube plug slightly when inserting it into the air |
tube connectors to avoid the air tube disconnecting during use. |
• To completely isolate the device from the power source unplug the plug from the |
power source. |
Keep these instructions for future reference.
EN
TM
Compressor Nebulizer
Model NE-C28P
Instruction Manual
| Compressed Air |
|
|
Manufacturer | OMRON HEALTHCARE Co., Ltd. |
| 53, Kunotsubo, |
|
|
OMRON HEALTHCARE EUROPE B.V. | |
| Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS |
| |
|
|
Production facility | OMRON (DALIAN) CO., LTD. |
| Dalian, CHINA |
|
|
Subsidiary | OMRON HEALTHCARE UK LTD. |
| Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, U.K. |
| OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH |
| |
| |
| OMRON SANTÉ FRANCE SAS |
| 14, rue de Lisbonne, 93561 |
| Made in China |
• | Do not block the ventilation slots. Never place the device where the ventilation slots |
| may be obstructed during operation. |
• | Store the parts in a clean location for avoiding infection. |
(Risk of electrical shock)
•Never plug or unplug the device with wet hands.
•The compressor and power cord are not waterproof. Do not spill water or other liquids on these parts. If liquid does spill on these parts, immediately unplug the power cord and wipe off the liquid with gauze or other absorbent material.
•Do not immerse the compressor in water or other liquid.
•Do not use or store the device in humid locations.
•Do not operate the device with a damaged power cord or plug.
•Keep the power cord away from heated surfaces.
(Cleaning and disinfecting)
Observe the rules below when cleaning or disinfecting parts. Failure to observe these rules may result in damage, inefficient nebulization or infection. For the instructions, refer to “Cleaning and disinfecting” section.
•Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before using them when as follows:
-the first time after purchase
-if the device has not been used for a long period of time
-if more than one person uses the same device
•Be sure to wash or wipe the parts after use, and ensure that they are thoroughly disinfected and dried, and stored in a clean location.
•Do not leave the cleaning solution in the parts.
Caution: | Indicates a potentially hazardous situation which if not avoided, may result in |
minor or moderate injury, or physical damage. |
(Usage)
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the
This medical device manufactured by OMRON HEALTHCARE conforms to this
Nevertheless, special precautions need to be observed:
•Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromagnetic fields, near the medical device. This may result in incorrect operation of the unit and create a potentially unsafe situation. Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case the distance is shorter.
Further documentation in accordance with
Documentation is also available at
Intended use
Medical Purpose
This product is intended to be used for inhaling medication for respiratory disorders.
Intended User
•Legally certified medical experts, such as doctor, nurse and therapist.
•Caregiver or patient under the guidance of qualified medical experts for home treatment.
•The user should also be capable of understanding general operation of
Intended Patients
This product should not be used on patients, who are unconscious or are not breathing spontaneously.
Know your unit
Compressor (Main Unit) x 1
Nebulizer Kit Holder
Ventilation Slots
Air Filter Cover
Air Tube Connector
Power Switch
Rear View
| Ventilation Slots |
Power Cord | Power Plug |
|
• Provide close supervision when this device is used by, on, or near children or invalids. |
• Make sure that the parts are attached correctly. |
• Make sure that the air filter is correctly attached. |
• Make sure that the air filter is clean. If the air filter has been used for more than |
60 days, replace it with a new one. |
• Do not tilt the nebulizer kit at an angle of greater than 45 degrees in all directions or |
shake it while in use. |
• Do not use or store the device while the air tube is creased. |
• Use only original nebulizing parts, air tube, air filter and air filter cover. |
• Do not add more than 7 ml of medication to the medication tank. |
• Do not carry or leave the nebulizer kit while the medication tank contains medication. |
• Do not leave the device unattended with infants or persons who cannot express their |
consent. |
• Do not subject the device or the parts to any strong shocks such as dropping the device |
on the floor. |
• Do not distort the vaporiser head, or jab the nozzle of the medication tank with a pin |
or any sharp object. |
• Do not insert fingers or objects inside the compressor. |
• Do not disassemble or attempt to repair the compressor or power cord. |
• Do not leave the device or its parts in extreme hot or cold temperature, or under direct |
sunlight. |
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it should not be disposed of, with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this product from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can return this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal.
Optional Medical Accessories
Environment
This product is intended for use in a medical facility, such as Hospital, clinic and doctor’s office, and in a room of general household.
Durable period
Durable periods are as follows, provided the product is used to nebulize saline 3 times a day for 10 minutes each time at room
temperature (23°C).
Durable period may vary depending on usage environment.
Compressor (Main unit) | 5 years |
Nebulizer Kit | 1 year |
Mouthpiece | 1 year |
Nosepiece | 1 year |
Air Tube (PVC, 200cm) | 1 year |
Air Filter | 60 days |
Mask (PVC)/(SEBS) | 1 year |
Precautions for use
Warnings and cautions described in the instruction manual should be observed.
Nebulizer Kit x 1 | Mouthpiece x 1 | Nosepiece x 1 |
Cap
Inhalation Air Inlet
Inhalation Top
Adult Mask (PVC) x 1 | Child Mask (PVC) x 1 |
Vaporiser Head
Nozzle
Medication Tank
Air Tube Connector
Air Tube | Spare Air Filters x 5 | Carrying Bag x 1 |
(PVC, 200cm) x 1 |
| Instruction Manual |
|
| |
|
| Warranty card |
• Do not block the air filter cover. |
• Do not use the device while sleeping or if drowsy. |
• Approved for human use only. |
• The compressor may become hot during operation. |
• Do not touch the compressor for other than necessary operation such as turning off |
the power while nebulizing. |
• To avoid the medication residue on the face, be sure to wipe the face after removing |
the mask. |
• To avoid injury to the nose mucosa, do not squeeze the nosepiece into the back of |
the nose. |
• Do not block the slit between the cap and the inhalation air inlet. |
• Make sure that the vaporiser head is correctly installed before use. |
• Do not use a damaged nebulizer kit, mouthpiece or nosepiece. |
(Risk of electrical shock)
•Always unplug the device from the power outlet after use and before cleaning.
•Plug the device into the appropriate voltage outlet. Do not overload power outlets or use extension cords.
•Do not misuse the power cord.
•Do not wind the power cord around the compressor.
•Changes or modifications not approved by OMRON HEALTHCARE will void the user warranty.
(Cleaning and disinfecting)
Observe the rules below when cleaning or disinfecting parts. Failure to observe these rules may result in damage, inefficient nebulization or infection. For the instructions, refer to “Cleaning and disinfecting” section.
•Do not use a microwave oven, dish dryer or hair dryer to dry the device or the parts.
•Do not use an autoclave, EOG gas sterilizer or low temperature plasma sterilizer.
•When disinfecting parts by boiling, make sure that the container does not boil dry. Otherwise it may also result in fire.
(within the scope of EC Medical Device Directive 93/42/EEC)
Name | Model | Order No. | |
V.V.T. Nebulizer Kit | |||
Mouthpiece Set | |||
|
| ||
|
|
| |
Adult Mask (PVC) | |||
|
|
| |
Child Mask (PVC) | |||
|
|
| |
Adult Mask (SEBS) | |||
|
|
| |
Child Mask (SEBS) Set | |||
|
|
| |
Mouthpiece | |||
|
|
| |
Nosepiece | |||
|
|
| |
Air Tube (PVC, 200cm) | |||
|
|
| |
Air Tube (Silicone, 100cm) | |||
|
|
|
Other Optional/Replacement Parts
Name | Model | Order No. |
Air Filters x 5 | ||
|
|
|