USING THE DEVICE

5Turn the power switch to the off ( ) position.

The compressor turns off and nebulization stops.

6Disconnect the air tube from the nebulizer kit.

Hold the air tube plug and gently pull down to disconnect the air tube plug from the air tube connector on the bottom of the nebulizer kit.

7Check the air tube. No condensation or moisture should remain in the air tube.

WARNING

Do not store the air tube with moisture or medication remaining in the air tube. This could result in infection as a result of bacteria.

If any condensation or moisture remains in the air tube, remove the moisture from the air tube. Follow the directions below:

1.Make sure the air tube is still connected to the air tube connector on the upper side of the compressor.

2.Press the power switch on. The compressor will start and pump air through the air tube to expel the moisture.

3.Press the power switch again to turn the compressor off.

8Disconnect the air tube from the compressor. Hold the air tube plug and gently pull the air tube plug off the air tube connector on the upper side of the compressor.

9Unplug the AC adapter from the

electrical outlet.

CAUTION

• Unplug the AC adapter from the electrical outlet after using the device.

• Remove the AC adapter from the device after use.

Do not pull the power cord of AC adapter strongly.