Guidance and Manufacturer's Declaration

Electromagnetic immunity IEC60601-1-2

Immunity test

IEC 60601

Compliance

Electromagnetic environment –

Test level

level

guidance

 

Electrostatic

±6 kV

±6 kV

Floor should be wood, concrete, or

ceramic tile. If floors are covered

discharge (ESD)

contact

contact

with synthetic material, the relative

IEC 61000-4-2

±8 kV air

±8 kV air

humidity should be at least 30 %.

 

 

 

 

±2 kV

 

 

Electrical fast

for power

±2 kV

Mains power quality should be that of

supply lines

for power

transient/burst

a typical commercial and/or hospital

±1 kV for

supply lines

IEC 61000-4-4

environment.

input/output

*1)

 

 

 

lines

 

 

 

±1 kV line

±1 kV line

Mains power quality should be that of

Surge

to line

to line

a typical commercial and/or hospital

IEC 61000-4-5

±2 kV line

±2 kV line

environment.

 

to earth

to earth

 

 

 

<5 % UT

<5 % UT

 

 

(>95 % dip

(>95 % dip

 

 

in UT) for

in UT) for

 

 

0.5 cycle

0.5 cycle

Mains power quality should be that of

 

40 % UT

40 % UT

 

a typical commercial and/or hospital

 

(60 % dip

(60 % dip

Voltage dips, short

environment.

in UT) for 5

in UT) for 5

interruptions and

If the user of the NE-C801 requires

cycles

cycles

voltage variations on

continued operation during power

70 % UT

70 % UT

power supply

mains interruption, it is recommended

(30 % dip in

(30 % dip in

IEC 61000-4-11

that the NE-C801 be powered from

UT) for 25

UT) for 25

 

an uninterruptible power supply or

 

cycles

cycles

 

battery.

 

<5 % UT

<5 % UT

 

 

 

(95 % dip

(95 % dip

 

 

in UT) for 5

in UT) for 5

 

 

sec.

sec.

 

Power frequency

 

 

Power frequency magnetic fields

(50/ 60 Hz) magnetic

3 A/m

3 A/m

should be at levels characteristic

field

of a typical location in a typical

 

 

IEC 61000-4-8

 

 

commercial or hospital environment.

Note: UT is the A.C. mains voltage prior to application of the test level.

*1) The test of input/output lines is not applicable since they are shorter than 3.0m.

28

Page 28
Image 28