M5-I
Hints |
| • Before measuring, stretch your back and sit up straight, | |||||
| breathe normally | ||||||
|
|
| |||||
|
|
| arms and entire body. | ||||
|
|
| • Lightly bend and find support for your elbow ( e.g. on the | ||||
|
|
| table ). | ||||
|
|
| • Do not move, talk or touch the device during measure- | ||||
|
|
| ment. | ||||
|
|
| • Stress influences your blood pressure. Relax before | ||||
|
|
| measurement. | ||||
|
|
| • To compare results measurements must always be | ||||
|
|
| taken from the same arm. | ||||
|
|
| • Do not place the cuff over thick clothes and do not roll up | ||||
|
|
| your sleeve if it is too tight. | ||||
|
|
| • Align the cuff with the height of your heart (breast | ||||
|
|
| height). If necessary use a pillow or cushion. | ||||
|
|
| • Coldness raises your blood pressure. Recommended is | ||||
|
|
| measuring at room temperature ( ±20°C). | ||||
|
|
| • Relax before measurement. If necessary go to the bath- | ||||
|
|
| room before measurement. | ||||
|
|
| • Do not measure directly after bathing. | ||||
|
|
| • When measuring in a moving vehicle the measurement | ||||
|
|
| can be influenced. | ||||
|
|
| • If you wish to interrupt measurement prematurely, press | ||||
|
|
| the ( |
| ) button. Inflation is stopped and the air is auto- | ||
|
|
| |||||
|
|
| matically released from the cuff. | ||||
|
|
| • If you wish to repeat the measurement, follow the same | ||||
|
|
| procedure as before starting at step 4 (Inflating the cuff). | ||||
|
|
| • If an error has occurred during measurement | E | is | ||
|
|
| displayed. Please refer to the next chapter. | ||||
|
|
| • The monitor should be switched off after measurement | ||||
|
|
| to avoid running down the batteries unnecessarily. If you | ||||
|
|
| forget to do so, the monitor switches itself off automati- | ||||
|
|
| cally after 5 minutes. | ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| English |
|
| Failure, causes and rectification | ||||||||||||||
| Failure and their possible causes |
| Rectification | ||||||||||||
| |||||||||||||||
|
|
|
|
|
|
|
|
|
|
| |||||
| Incorrect measurement is indicated by |
|
|
|
|
|
|
|
| ||||||
| E |
|
|
|
|
|
| ||||||||
1. | A correct reading could not be obtained |
| 1. | Repeat the measurement keeping | |||||||||||
|
|
| because measurement was disturbed by |
|
| perfectly still. Do not move your arm and | |||||||||
|
|
| movement of the body. |
|
| do not speak. | |||||||||
2. | The cuff is not fitted correctly. |
| 2. | Check that the cuff is correctly fitted. Then | |||||||||||
|
|
|
|
|
|
|
|
|
|
|
| repeat the measurement. | |||
3. | Your clothing has constricted your blood flow. |
| 3. | Remove the item of clothing which cau- | |||||||||||
|
|
|
|
|
|
|
|
|
|
|
| sed the constriction. | |||
4. | There is still air in the cuff when the monitor |
| 4. | The unit may be defective. Please have it | |||||||||||
|
|
| is switched on. |
|
| examined and, if necessary, repaired by | |||||||||
|
|
|
|
|
|
|
|
|
|
|
| OMRON Customer Services. | |||
|
|
|
|
|
|
|
|
| |||||||
| The start button was pressed before the |
| Wait for the ( |
| ) symbol indicating readi- | ||||||||||
|
|
| |||||||||||||
( |
|
| ) symbol was displayed |
| ness before |
| press the start button. | ||||||||
|
|
| you | ||||||||||||
|
|
|
|
|
|
|
|
| |||||||
| The blood pressure values displayed are ex- |
| Refer to the instructions and then repeat the | ||||||||||||
| tremely low or high, or they are implausible. |
| measurement. | ||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| |||
| The display does not light up when the ( |
| ) |
|
|
|
|
|
| ||||||
|
|
|
|
|
|
| |||||||||
| button is pressed. |
|
|
|
|
|
| ||||||||
1. | The batteries are exhausted. |
| 1. | Check the batteries and, if necessary, fit 4 | |||||||||||
|
|
|
|
|
|
|
|
|
|
|
| new batteries. | |||
2. | The +/- poles of the batteries have been |
| 2. | Reinsert the batteries with the +/- poles | |||||||||||
|
|
| reversed. |
|
| the right way round. | |||||||||
3. | The battery contacts are dirty. |
| 3. | Clean the battery contacts with a dry cloth. | |||||||||||
|
|
|
| The batteries are weak or exhausted. |
| Fit new batteries. | |||||||||
|
|
|
| ||||||||||||
|
|
|
|
|
|
|
|
| |||||||
| The cuff pressure does not rise although |
| Check that the air tube is properly connec- | ||||||||||||
| the pump motor can be heard. |
| ted to the monitor. Push the tube connector | ||||||||||||
|
|
|
|
|
|
|
|
|
|
| firmly into the socket. | ||||
|
|
|
|
|
| Contact Omron Customer Service. | |||||||||
|
|
| Err | problem with mem. function |
| ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|