CLEANING AFTER EACH USE

Following the cleaning instructions after each use will prevent any remaining medication in the bottle from drying resulting in the device not nebulizing effectively and will help prevent infections.

WARNING

Wash the nebulizer parts after each use. Dry the parts immediately after washing.

1.Remove the inhalation accessory (mouthpiece, mask, or nosepiece) from the nebulizer kit.

2.Disconnect the air tubing from the nebulizer.

3.Gently twist the inhalation top counterclockwise and lift to separate the nebulizer into two sections.

4.Remove the baffle.

5.Discard remaining medication in the medication cup.

WARNING

3 5 ax

m

Always dispose of any remaining medication in the medication cup after each use. Use fresh medication each time you use the device.

6.Rinse all the parts of the nebulizer kit and the mouthpiece or nosepiece with hot tap water. You may also wash these parts in a dishwasher. Remove the parts after the wash and rinse cycle is completed.

DO NOT wash the mask in a dishwasher.

CAUTION

Place the nebulizer pieces in a basket recommended for dishwasher use. Place the basket on the top shelf to avoid any damage to the nebulizer kit.

7.Rinse the mask with hot tap water.

8.Hand dry or air dry in a clean environment using a soft, clean, lint-free cloth.

9.Assemble the nebulizer and store the nebulizer kit in a dry, sealed bag.

53

CAUTION

The nebulizer kit should be replaced every 6 months.