| 6 Error: causes and rectifications | |||||||
|
|
|
|
|
|
|
|
|
| Indication | Error | Cause | Rectification | ||||
|
|
|
|
|
|
|
| |
|
|
|
|
| Correct | Cuff pressure | Wait for at least three minutes to repeat | |
|
|
|
|
| reading | low | measurement. | |
|
|
|
|
| not |
| Inflate the cuff 40 mmHg higher than | |
|
|
|
|
|
| |||
|
|
|
|
| possible | Movement | previous value. | |
|
|
|
|
|
|
| Follow instructions, repeat | |
|
|
|
|
|
|
| ||
|
|
|
|
|
| Very weak | measurement. | |
|
|
|
|
|
| artery pulse |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
| Cuff pressure | Wait for at least three minutes to repeat | |
|
|
|
|
|
| too high | measurement. | |
|
|
|
|
|
|
| Inflate the cuff 30 mmHg lower than | |
|
|
|
|
|
|
| previous value. | |
|
|
|
|
|
|
| Follow instructions, repeat | |
|
|
|
|
|
|
| measurement. | |
|
|
|
|
|
|
|
| |
|
|
|
|
| Low | Batteries weak/ | Fit four new identical 1.5V | |
|
|
|
|
| ||||
|
|
|
|
| voltage | exhausted | alkaline batteries type AA | |
|
|
|
|
|
|
| LR6 | |
| Display empty |
|
|
|
| |||
|
|
|
|
|
|
|
|
|
7 Maintenance and spare parts
•Use a soft, slightly moistened cloth to clean the OMRON MX3 Plus. Caution! Do not use petrol, thinners or similar solvents!
•Use a soft, moistened cloth and soap to clean the cuff.
Caution! Do not wash the cuff! Caution! Do not use petrol, thinners or similar solvents!
•Do not carry out repairs of any kind yourself. If a defect occurs consult your
OMRON retail outlet or distributor as mentioned on the packaging.
Calibration
•It is recommended to have the OMRON MX3 Plus inspected every two years to ensure correct function and accuracy.
Consult your OMRON retail outlet or OMRON distributor as mentioned on the packaging.
Replace batteries
• Remove the battery out of the battery
8