Using the Imaging Flip Mirror

Figure 5.1. Lever up directs the light path to the CCD camera. Figure 5.2. Lever down directs the light path to the eyepiece.

Using the Imaging Flip Mirror

Keep the flip mirror lever in the “up” position (Figure 5.1) to let the telescope focus on the camera. Flip the mirror to the “down” posi- tion to let the telescope focus on the eyepiece (Figure 5.2).

Lever tension adjustment screws

Figure 6. (above) The two silver socket head screws located above the lever adjust the lever tension.

Figure 7. (right) The socket head set screw located beneath the flip mirror body adjusts the mirror stop angle.

Flip Mirror stop angle adjustment set screw

With the flip mirror lever in the “up” position, focus your CCD cam- era using your telescope’s focuser. Then, flip the mirror to the “down” position to switch the light path to your eyepiece. Focus the eyepiece using the helical focus mechanism, making sure not to disrupt the telescope’s focuser. When your eyepiece is focused, lock the helical focus mechanism by tightening the focus lock thumbscrew.

Other Adjustments

The flip mirror lever tension and stop angle are factory set. However, if needed, adjustments can be made to change the lever tension and stop angle.

Adjusting the Lever Tension

Locate the two silver socket head set screws (one on each side) located above the lever (Figure 6).

Using the provided 1.5mm hex key, loosen or tighten the set screws as needed to reach the desired lever tension.

Adjusting the Flip Mirror Stop Angle

Check if any adjustment is necessary first. The Imaging Flip Mirror stop angle is factory-set and typically does not require adjustment. Center the image subject in your CCD camera. Then flip the mirror down to bring the light path to the eyepiece. The subject should be centered in the eyepiece. If the subject is not centered in the eyepiece as is was in the CCD camera, you need to adjust the mirror stop angle.

To adjust the flip mirror stop angle:

1.Locate the socket head set screw on the bottom of the flip mirror body (Figure 7).

2.If the subject is above the center of the field of view (when facing the rear of the flip mirror), turn the set screw clockwise

(using the provided 2.5mm hex key). If the subject is below the center of the field of view, turn the set screw counter-clockwise.

3.Check your adjustment by comparing the CCD camera view with the eyepiece view. The subject should be centered in both views.

Care and Storage

The mirror surface should only be cleaned on a minimal bases. Small dust particles will not effect the optical performance.You can occasionally remove dust with compressed air or a blower bulb. If you need to clean the mirror surface, first unthread the 1.25” eye- piece holder from the flip mirror body to allow easy access to the mirror. Use a cotton Q-tip with a drop of isopropyl alcohol or lens cleaning solution. Lightly drag the Q-tip across the surface of the mirror. Replace the Q-tip frequently while cleaning to ensure you do not smear particles across the mirror. Use caution, as scratch- ing or stripping the delicate aluminum surface can easily occur.

To avoid excessive dust buildup, be sure to attach the dust caps prior to storing, and use a foam-lined accessory case. The foam- lined box your Imaging Flip Mirror arrives in also acts as a con- venient storage container. Store in a dry place away from direct sunlight.

One-Year Limited Warranty

This Orion Imaging Flip Mirror is warranted against defects in materials or workmanship for a period of one year from the date of purchase. This warran- ty is for the benefit of the original retail purchaser only. During this warranty period Orion Telescopes & Binoculars will repair or replace, at Orion’s option, any warranted instrument that proves to be defective, provided it is returned postage paid to: Orion Warranty Repair, 89 Hangar Way, Watsonville, CA 95076. If the product is not registered, proof of purchase (such as a copy of the original invoice) is required.

This warranty does not apply if, in Orion’s judgment, the instrument has been abused, mishandled, or modified, nor does it apply to normal wear and tear. This warranty gives you specific legal rights, and you may also have other rights, which vary from state to state. For further warranty service information, contact: Customer Service Department, Orion Telescopes & Binoculars, 89 Hangar Way, Watsonville, CA 95076; (800)676-1343.

 

 

 

 

 

Customer Support (800)676-1343

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

E-mail: support@telescope.com

 

 

 

 

 

 

 

 

 

 

 

Providing Exceptional Consumer Optical Products Since 1975

Corporate Offices (831)763-7000

 

 

 

 

OrionTelescopes.com
89 Hangar Way, Watsonville CA 95076

 

 

 

 

 

 

 

 

 

 

 

 

Page 2
Image 2
Orion 5523 manual Using the Imaging Flip Mirror, Other Adjustments, One-YearLimited Warranty, Adjusting the Lever Tension