Polaroid MicroCam manual Microscope information, Magnification

Models: MicroCam

1 16
Download 16 pages 59.44 Kb
Page 11
Image 11

English 9

Xenon lighting

If using a Xenon (daylight) microscope illuminator with color film, set the camera for Type 331 black & white film, to remove the built-in filter. Then press the black arrow ▼ to decrease the exposure by × .66 to compensate for the difference in film speed.

Dark-field illumination

The camera's automatic exposure system may produce incorrectly exposed pictures when dark-field illumination is used. The built-in exposure meter can be used to determine the correct exposure as follows:

Set the microscope for bright-field illumination.

Set the camera for Automatic mode, and take a meter reading as described above.

Re-set the microscope for dark-field illumination, and the camera for manual mode.

Use the exposure time indicated by the camera's meter to make a manual exposure.

Microscope information

Because of the great variety of microscopes that are compatible with the MicroCam, this booklet cannot provide specific details for all equipment. You should refer to the instructional material provided with your microscope, or contact the supplier for more information.

One of the most important microscope adjustments for photography is the illumination. The microscope illumination must meet specific criteria: The beam of light must be correctly concentrated and adjusted to provide even illumination over the entire picture area, as the camera cannot compensate for uneven lighting.

Please refer to the “universal” microscope illustration (17) for the location of microscope controls.

Note: Some older microscopes utilize optics in the eyepiece to correct for chromatic aberration. An adapter will be required to produce optimum resolution with the MicroCam. (Contact Polaroid Technical Assistance for information.)

Magnification

The final magnification of the print is approximately ten times the magnifi- cation power of the microscope's objective lens. If the exact magnification must be verified, include a scale in the photograph.

Page 11
Image 11
Polaroid MicroCam manual Microscope information, Magnification