ReliOn K 144-732-000 specifications What is a NormalTemperature?, Taking an Oral Temperature

Models: K 144-732-000

1 2
Download 2 pages 19.42 Kb
Page 1
Image 1

FAST 10-Second

Digital Thermometer

Thank you for purchasing a ReliOn® Digital Thermometer. Please read the following instructions to achieve the most accurate temperatures and safe operation. With proper care and use, your thermometer will provide you with years of reliable readings.

What is a “Normal”Temperature?

Although the generally accepted “normal” temperature reading is 98.6oF, tempera- ture readings can vary from 97oF to 99oF and still may be considered “normal”. Variations in temperature can be attributed to activities such as exercise, smoking, eating and drinking. Even time of day may influence your temperature. For exam- ple, your temperature is lower in the morning than in the afternoon. Other variations may be due to the location of the temperature reading. While the oral temperature follows the guidelines previously described, a rectal temperature is generally 1.0oF higher. Conversely, an axillary (under the arm) temperature will be 1.0oF lower.

Taking an Oral Temperature

Approximate measurement time: 15 seconds

1.Using the diagram on the right, place the thermometer tip tip under the tongue as near as possible to the heat pocket. The diagram also shows the temperature variables by not placing the tip in the heat pocket area.

NOTE: Drinking hot or cold fluids, exercis- ing, smoking or performing other activities may raise or lower your temperature.

Therefore, it is important to relax for approximately 5 minutes with your mouth closed prior to taking a reading

HEAT

HEAT

POCKET 98.8oF

98.8oF POCKET

98.4oF

98.4oF

 

98.1oF

97.9oF

97.9oF

 

97.7oF

General Care

1.The thermometer should only be used under the supervision of an adult.

2.Do not walk, run or talk during temperature taking.

3.Clean the thermometer before and after each use.

4.Store the unit in the protective case when not in use.

5.Do not bite the thermometer tip.

6.Do not store the unit where it will be exposed to direct sunlight, dust or humidity. Avoid extreme temperatures.

7.Dropping or subjecting your thermometer to strong shocks should be avoided.

8.Do not attempt to disassemble the unit, except to replace the battery.

9.If ERR displays and measurement is not attainable, the unit must be replaced.

NOTE: Performance of the device may be degraded if: operated outside of stated temperature and humidity range; stored outside of stated temperature and humidity range; thermometer undergoes mechanical shock (drop); patient temperature is below ambient temperature.

Changing the Battery

LIFETIME LIMITED WARRANTY

This ReliOn® Digital Thermometer is guaranteed to be free from manufacturing defects for the life of the original registered purchaser under normal use. This limited warranty does not apply to use in any clinical or commercial applications. If the unit fails to operate during the lifetime of the original registered purchaser, return it postage prepaid to: Repair Department, 1931 Norman Drive, Waukegan, IL 60085. If ReliOn® determines that the unit failed to operate due to a manufacturing defect, the unit will be repaired or replaced at the option of ReliOn®. Repair or replacement of the unit is the sole remedy under this limited warranty. This warranty gives you specific legal rights, which vary from state to state. As a condition of this warranty, the enclosed warranty registration card must be completed and sent to us within 10 days of the purchase date.

This Limited Warranty constitutes ReliOn®’s only responsibility and obligation to repair and/or replace materials or components. We make no other express or implied warranties, arising by operation of law or otherwise, or any warranty of merchantability or fitness for a particular use or purpose whether or not the use or purpose has been disclosed to ReliOn® in specifications, drawings or otherwise, and whether or not ReliOn®’s products are specifically designed and/or manufactured by ReliOn® for the buyer’s use or purposes, except for the limited warranty stated above. ReliOn® will not be responsible for any indirect, incidental, special, consequential, or punitive damages or other loss, including, but not limited to, damage to or loss of other property or equipment and personal injuries, whether to purchaser or others. ReliOn® shall in no event be liable to the purchaser for any amount in excess of the cost of repair and/or replacement of the unit

Features of your ReliOn® Digital Thermometer

DIGITAL DISPLAY WINDOW BATTERY COVER

THERMOMETER TIP

ON/OFF BUTTON

Cleaning Instructions

Before and after each use, clean the thermometer with a soft cloth and 70% isopropyl alcohol diluted with water, or wash with COLD soapy water. An intact unit can be submersed in water or alcohol. However, do not soak for an extended period of time.

1.The unit must never be sterilized by boiling, gas or steam autoclaves.

2.The use of any other solution or cleaning method beyond the above mentioned may result in reliability problems with the thermometer and void your warranty.

How to Take Your Temperature

1.Clean and disinfect the thermometer according the instructions provided.

2.Apply a disposable thermometer tip cover to the clean, dry thermometer.

3.Determine whether the measurement site will be oral, rectal or underarm. NOTE: Measurement time may vary based on the selected site.

4.Press the ON/OFF button (the unit will beep) to activate the thermometer.

This is the LCD display test, which insures all parts of the dis- play are functioning properly. Continue to hold the button down and in approximately 2 seconds the display will show the last recorded temperature as follows.

This display shows the last temperature reading. Within approximately 2 seconds of releasing the button or if no reading is stored in memory, the display will read as follows.

This display denotes normal operation. “Err” denotes a malfunction.

2.The current temperature is shown and the oF will continue to flash during the reading.

3.When the peak temperature has been reached, the oF will stop flashing and the thermometer will sound four sets of five quick beeps.

4.To turn the unit off, press the ON/OFF button. The unit will automatically shut off after approximately nine minutes.

NOTE: The mouth must remain closed during the measurement for the most accu- rate reading. Opening the mouth could result in an extended measurement time and may affect the reading.

Taking an Axillary Temperature (under the arm)

Approximate measurement time: 20 seconds

1.Wipe the underarm with a dry towel.

2.Place the thermometer tip under the arm so the tip is touching the skin with the thermometer perpendicular to the body.

3.Position the arm across the chest so the thermometer tip tip is well covered by the arm. This also insures the thermometer tip is not affected by the room’s air.

4.When the peak temperature has been reached, the oF will stop flashing and the thermometer will sound four sets of five quick beeps.

5.To turn the unit off, press the ON/OFF button. The unit will automatically shut off

after approximately nine minutes.

NOTE: Axillary temperatures are generally 1.0oF lower than oral readings.

Taking a Rectal Temperature

Approximate measurement time: 10 seconds

NOTE: Rectal temperatures are an appropriate and reliable method for infants and small children. Otherwise, this method should only be used when it is impossible or impractical to take an oral or axillary temperature.

1.After applying the disposable thermometer tip cover, apply a water-soluble lubri- cant to the tip.

2.Gently insert the thermometer tip a MAXIMUM of 1/2” into the rectum.

3.When the peak temperature has been reached, the oF will stop flashing and the thermometer will sound four sets of five quick beeps.

4.To turn the unit off, press the ON/OFF button. The unit will automatically power

down after approximately ten minutes.

NOTE: Rectal temperatures are generally 1.0oF higher than oral readings.

Using Disposable Thermometer Tip Covers

The battery in your digital thermometer needs to be replaced when the “” sym- bol appears in the lower right of the display. Replace the battery with a 1.55V, AG3 alkaline or SR41 silver oxide type or equivalent.

To replace the battery,

1.Slide the battery cap away from the unit.

2.Using a non-metal pointed tool, carefully remove the battery and insert the new battery with the negative (–) side facing the back side of the unit.

3.Carefully slide the battery cap back into place, making sure that the cap fits snugly to protect the battery com-

partment from moisture.

NOTE: Please properly dispose the battery away from small children and heat.

Specifications

DISPLAY:

Liquid Crystal Display

 

 

DISPLAY RESOLUTION:

4 Digits (1/10TH of a degree increments)

 

 

MEASUREMENT RANGE:

90.0˚F to 111.9˚F

 

 

ACCURACY:

±0.2˚F

 

 

AVERAGE MEASUREMENT TIME:

Oral 15.2 seconds. Rectal 10.2 seconds. Axillary 20.8 seconds

 

 

TONE:

Thermometer beeps when activated and deactivated.

 

A series of beeps signal when peak temperature is reached.

 

 

MEMORY:

Displays last measured temperature

 

 

BATTERY:

One, 1.55V, AG3 alkaline

 

 

BATTERY LIFE:

Approximately 4,000 measurements

 

 

AUTOMATIC SHUT-OFF:

After approximately 9 minutes

 

 

AMBIENT OPERATING TEMPERATURE:

50.0˚F to 113.0˚F

 

 

DIMENSIONS:

5-3/16” x 11/16” x 3/8”

 

 

WEIGHT:

0.34 ounces

 

 

Specifications are subject to change without notice.

Manufactured for:

MABIS Healthcare Inc.

1931 Norman Drive

Waukegan, IL 60085 USA

www.mabis.net

CUT OUT AND MAIL TO:

WARRANTY REGISTRATION DEPARTMENT

1931 NORMAN DRIVE

WAUKEGAN, IL 60085

THERMOMETER WARRANTY REGISTRATION

(MUST BE COMPLETED WITHIN 10 DAYS OF PURCHASE)

NAME:

ADDRESS:

CITY:

 

 

 

STATE:

 

ZIP:

 

PURCHASED FROM:

 

 

 

 

 

 

CITY:

 

 

STATE:

 

ZIP:

 

DATE PURCHASED:

 

 

 

 

 

 

 

MODEL PURCHASED:

 

 

 

 

 

144-732-000

 

 

 

 

USER INFORMATION (OPTIONAL)

 

 

 

 

 

MALE

FEMALE

AGE:

 

 

AGE OF YOUNGEST FAMILY MEMBER

UNDER 6 MONTHS

6 MONTHS - 1 YEAR

1 - 3 YEARS

3 - 5 YEARS

OLDER

PROBE COVER ORDER FORM

CUT OUT AND MAIL TO:

MABIS HEALTHCARE INC. 1931 NORMAN DRIVE WAUKEGAN, IL 60085

Please send packages of disposable probe covers #15-618-000 (100 probes per package). Enclosed is $5.00 per package plus $3.00 for shipping and handling for each order.

PAYMENT METHOD (do not send cash): Check Money Order

Please ship to:

NAME:

ADDRESS:

CITY:

 

 

STATE:

 

ZIP:

 

PHONE: (

)

 

 

 

 

 

 

 

 

 

 

 

 

NOTE: This order form is for shipments within the United States and Canada.

The Lo with the flashing oF indicates the thermometer is ready to take a temperature reading.

It is recommended to use the enclosed disposable thermometer tip covers to pre- vent cross contamination. ReliOn® thermometer tip covers can be purchased from your local Wal-Mart retailer or with the order form provided.

Meets ASTM Standard E1112

Made in China

ReliOn® is a trademark under license by MABIS Healthcare Inc.

#144-ACT2030 05/04 ©2004 ReliOn®