ResMed Elise 150 ELISEE 150 TECHNICAL SPECIFICATIONS, Paediatric, Adult, Adjustments, Parameters

Models: Elise 150

1 4
Download 4 pages 56.5 Kb
Page 4
Image 4
ELISEE 150 TECHNICAL SPECIFICATIONS

ELISEE 150 TECHNICAL SPECIFICATIONS

 

Paediatric

 

Adult

 

Adjustments

 

 

 

 

 

 

 

PARAMETERS

 

 

 

 

 

 

Pressure slope

 

1, 2, 3, 4

 

1, 2, 3, 4

 

Inspiration pressure (cm H2O)

 

3–40 in NIV

 

5–60

 

1 cm H2O

 

 

3–60 in Invasive

 

 

 

 

 

 

 

 

Pressure support (cm H2O)

 

3–40 in NIV

 

5–60

 

1 cm H2O

 

 

3–60 in Invasive

 

 

 

 

 

 

 

 

PEEP (cm H2O)

 

0–20

 

0–25

 

1 cm H2O

CPAP (cm H2O)

 

3–20

 

3–25

 

1 cm H2O

Tidal volume (mL)

 

50–500

 

300–2500

 

10 mL (50–1000)

 

 

 

 

100 mL (1000–2500)

 

 

 

 

 

 

Security tidal volume (mL)

 

NO / 50–500

 

NO / 300–2500

 

10 mL (50–1000)

 

 

 

 

100 mL (1000–2500)

 

 

 

 

 

 

Maximum flow setting (volume) (L/min)

 

5–40

 

10–120

 

1 L/min

Flow shape

 

(1–5) Square/Decreasing

 

(1–5) Square/Decreasing

 

I/E ratio (1/x)

 

1/0.4–1/9.9

 

1/0.4–1/9.9

 

1/0.1

Respiratory rate

 

2–80

 

2–50

 

1 bpm

Minimum breath (bpm)

 

NO / 3–80

 

NO / 6–50

 

1 bpm

Maximum inspiration time (seconds)

 

0.3–3.0

 

0.3–3.0

 

0.1 sec

Apnea time (seconds)

 

5–20

 

10–60

 

1 sec

TRIGGERS

 

 

 

 

 

 

Inspiration pressure trigger (cm H2O)

 

0.2–5.9 / NO in

 

0.2–5.9 / NO in

 

 

(Invasive ventilation only)

 

(A)CV and P(A)CV modes

 

(A)CV and P(A)CV modes

 

0.1 cm H2O

 

 

0.2–6.0 in PS.SV, PS.VT,

 

0.2–6.0 in PS.SV, PS.VT,

 

 

 

 

 

 

 

 

SIMV and PSIMV modes

 

SIMV and PSIMV modes

 

 

Inspiration flow trigger (L/min)

0.2–9.9 / NO in

 

0.2–9.9 / NO in

 

 

(Invasive ventilation only)

 

(A)CV and P(A)CV modes

 

(A)CV and P(A)CV modes

 

0.1 L/min

 

 

0.2–10.0 in PS.SV, PS.VT,

 

0.2–10.0 in PS.SV, PS.VT,

 

 

 

 

 

 

 

 

SIMV and PSIMV modes

 

SIMV and PSIMV modes

 

 

Inspiration trigger

 

AUTO / 1–5 in PS.SV,

 

AUTO / 1–5 in PS.SV,

 

 

(Noninvasive ventilation only)

 

PS.VT, SIMV and

 

PS.VT, SIMV and

 

 

PSIMV modes

 

PSIMV modes

 

 

AUTO / 1–5 / NO in

 

AUTO / 1–5 / NO in

 

 

 

 

(A)CV and P(A)CV modes

 

(A)CV and P(A)CV modes

 

 

Expiratory trigger

 

10–90 / AUTO in PS.SV, PS.

 

10–90 / AUTO in PS.SV, PS.

 

1 %

(% peak flow)

 

VT, SIMV and PSIMV modes

 

VT, SIMV and PSIMV modes

 

 

 

 

 

LUNG RECRUITMENT

 

 

 

 

 

 

Period (minutes)

 

NO / 1–60

 

NO / 1–60

 

1 min

Duration (seconds)

 

0.2–40

 

0.3–40

 

0.1 sec (0.2–3)

 

 

 

 

1 sec (3–40)

 

 

 

 

 

 

Pressure (cm H2O)

 

5–60

 

5–60

 

1 cm H2O

TECHNICAL DATA

 

 

 

 

 

 

External DC supply

 

 

 

 

 

12–28 VDC, maxi 15 amps

Internal battery

 

 

 

 

 

Li-Ion, 14.4 VDC, 6.3 AH

External battery

 

 

 

 

 

Li-Ion, 14.4 VDC, 6.3 AH

Power consumption

 

 

 

 

 

max 75 VA

External power

 

 

 

 

 

12–28 V

Internal battery capacity*

 

 

 

 

 

4–6 hrs

External battery capacity*

 

 

 

 

 

4–6 hrs

Size

 

 

 

26 cm x 24 cm x 13 cm (10.2” x 9.4” x 5.1”)

Mains supply

 

 

 

100 VAC/1 amp to 230 VAC / 0.6 amp – 50/60 Hz

Weight

 

 

 

4.45 kg (9.79 lb) with internal and external battery pack

* Tested under specific settings. Please refer to the User Guide.

Manufacturer: ResMed Paris 240 rue de la Motte 77550 Moissy-Cramayel France. Distributed by: ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia, ResMed (UK) Ltd 96 Milton Park Abingdon

Oxfordshire OX14 4RY UK. See www.resmed.com for other ResMed locations worldwide. Protected by patents: FR 0303538, FR 2766568, US 6164141. Other patents pending. Elisée is a registered trademark of ResMed Paris. Specifications may change without notice. ResMed Paris is a subsidiary of ResMed Inc. © 2009 ResMed Ltd. 1013308/1 09 01

Global leaders in sleep and respiratory medicine www.resmed.com

Page 4
Image 4
ResMed Elise 150 ELISEE 150 TECHNICAL SPECIFICATIONS, Paediatric, Adult, Adjustments, Parameters, Triggers, Technical Data