ResMed VPAP S manual Vpap S Features Summary, Technical Specifications, Product Codes

Models: VPAP S

1 2
Download 2 pages 23.72 Kb
Page 2
Image 2
VPAP S FEATURES SUMMARY

VPAP S FEATURES SUMMARY

Modes of operation

S (Spontaneous)

CPAP (with EPR)

Product features

Simplified standard setting menu

Vsync automatic leak management

TiControl (Ti max/min)

Adjustable breath trigger/cycle

24V DC input (via converter)

Compact S8design

Easy-Breathe motor

Heated humidification

• Optional, fully integrated

Data storage

Usage and treatment summary (day, week, month, six-month, year)

Usage profile

Summary data storage

Detailed data storage

Remote settings adjustment

Data transfer

Direct download to PC

ResLink via SmartMedia card

ResScan Data Card

Live oximetry with ResLink

Remote compliance monitoring via ResTraxx

TECHNICAL SPECIFICATIONS

Performance

Operating pressure range: 3 to 25 cm H2O

Maximum single fault pressure: 40 cm H2O

Dynamic pressure characteristics

S mode: IPAP: 4 to 25 cm H2O; EPAP: 3 to 25 cm H2O

CPAP mode: 4 to 20 cm H2O

Sound pressure level

<26 dB (tested in accordance with the requirements of ISO 17510-1:2002)

Data storage

Data includes usage, leak, pressure, AHI, AI, tidal volume, respiratory rate and minute ventilation

Via LCD menu: 365 sessions of summary data

Via ResLink: 365 sessions of summary data, 30 sessions of detailed data

Via ResScan Data Card: 180 sessions of summary data, two sessions of detailed data

Dimensions (L x W x H): 4.4” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm)

Weight: 2.9 lb (1.3 kg)

Power supply

Input range 100–240VAC, 50–60Hz, 40VA (typical power consumption),

<100VA (maximum power consumption), 12–24DVC via ResMed DC 24V/50W converter

Environmental conditions

Operating temperature: 41°F to 97°F (+5°C to +35°C)

Operating humidity: 10–95% non-condensing

Electromagnetic compatibility

Product complies with all applicable electromagnetic compatibility requirements (EMC) for residential, commercial and light industry environments according to IEC 60601-1-2.

Air filter

Two-layered, powder-bonded, polyester non-woven fiber

IEC 60601-1 classification

Class II (double insulated), Type CF

Note: The manufacturer reserves the right to change these specifi cations without notice.

PRODUCT CODES

VPAP S

26119

H4i with standard water chamber

26940

VPAP S + H4i

26120

S8

ResLink

22206

Oximeter Kit

1430050

S8

ResTraxx US

22204

S8

ResTraxx Canada

22221

ResMed Corp Poway, CA, USA +1 858 746 2400 or 1 800 424 0737 (toll free), ResMed Ltd Bella Vista, NSW, Australia +61 (2) 8884 1000 or 1 800 658 189 (toll free). Offices in Austria, Belgium, Brazil, China, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands, New Zealand, Norway, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, United Kingdom (see website for details). EPR, HumidAire, ResLink, S8, ResScan, ResTraxx, TiControl, and Vsync are trademarks of ResMed Ltd. H4i and VPAP are trademarks of ResMed Ltd and are registered in U.S. Patent and Trademark Office. ©2008 ResMed. 1011085/1 08 10

Global leaders in sleep and respiratory medicine www.resmed.com