English 13
Pressure accuracy Maximum static pressure variation at 10 cm H2O according to EN ISO 17510-1:2009
Standard air tubing SlimLine air tubing
Without H5i 9.89 cm H2O to 9.97 cm H2O 9.76 cm H2O to 9.87 cm H2O
With H5i 9.82 cm H2O to 9.98 cm H2O 9.78 cm H2O to 9.88 cm H2O
Maximum dynamic pressure variation according to EN ISO 17510-1:2009
Pressure (cm H2O) 10 BPM 15 BPM 20 BPM
VPAP ST and Standard air tubing without H5i / VPAP ST and Standard air
tubing with H5i
4 0.18 / 0.18 0.30 / 0.30 0.51 / 0.51
8 0.21 / 0.20 0.26 / 0.24 0.38 / 0.36
12 0.21 / 0.20 0.26 / 0.23 0.34 / 0.31
16 0.22 / 0.21 0.27 / 0.26 0.36 / 0.33
20 0.23 / 0.22 0.26 / 0.28 0.38 / 0.35
25 0.30 / 0.31 0.54 / 0.50 0.74 / 0.71
Pressure (cm H2O) 10 BPM 15 BPM 20 BPM
VPAP ST and SlimLine air tubing without H5i / VPAP ST and SlimLine air
tubing with H5i
4 0.22 / 0.20 0.28 / 0.29 0.47 / 0.53
8 0.23 / 0.19 0.32 / 0.29 0.41 / 0.42
12 0.22 / 0.21 0.35 / 0.29 0.41 / 0.45
16 0.22 / 0.23 0.41 / 0.33 0.44 / 0.50
20 0.24 / 0.27 0.37 / 0.34 0.48 / 0.50
25 0.31 / 0.31 0.50 / 0.54 0.78 / 0.84
The following symbols may appear on your product or packaging.
Caution; Read instructions before use; Protection against insertion of fingers and against
vertically dripping water; Type BF equipment; Class II equipment; Start/Stop;
Manufacturer; European RoHS; Batch code; Catalogue number; Serial number;