VI. Specifications

Sample Parameters:

Vapor only, oil and liquid free, 300 psig (2 MPa) Maximum

Detected Compounds:

R12, R134a, Air and unknown compounds

Sensor Technologies:

Non-Dispersive Infrared and Chemical Fuel Cell

Refrigerant Sample Size:

0.12 ounces (3.5 grams) per typical sample – weight of vapor loss

Power (Standard):

Powered by vehicle battery through clip-on adapter harness (10 to 14 VDC).

Power (Optional):

Powered by line power through plug-in wall adapter harness, 115 VAC unit (OPTIONAL) or 220 VAC unit (OPTIONAL)

Operational Temperature Range:

45-100oF (7-38oC)

Instrument Weight:

1.5 Lbs. (454 grams)

R12/R134a Purity Setpoint:

Pass if single refrigerant, R12 or R134a, is detected in concentrations equal to or greater than 98% by weight.

Unknown Refrigerant Purity Setpoint:

In the presence of detected unknown refrigerant types R12 or R134a is considered pure if detected in concentrations equal to or greater than 96% by weight.

Effect of Air Concentrations:

Air is not considered a contaminate and will be removed from the weight concentration determination of refrigerants. Air concentrations will only be reported for blends of R12 and R134a refrigerants.

SAE J1771:

If the refrigerant being tested is identified as contaminated, any visual percentages displayed of CFC-12 (R-12) or HFC-134a (R- 134a) outside the design certified value is informational and may not be accurate. SAE J1771 Certification Note: This equipment has not been designed certified for contaminate detection of R-124 or R-142B.