Sanyo MDF-793 Operating Instructions, Temperature setting, Remote alarm terminal, Key operated

Models: MDF-593 MDF-593AT MDF-793 MDF-793AT

1 30
Download 30 pages 58.12 Kb
Page 17
Image 17
OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS

Temperature setting

Table 1 shows the basic operation method. Perform key operations in the sequence indicated in the table. The example in the table is based on the assumption that the temperature is -70oC.

Note: The unit is set at the factory that the chamber temperature is -85C.

Table 1. Basic operation sequence (Example: Chamber temperature -70oC)

 

Description of operation

Key operated

Indication after operation

1

Turn on the power switch.

-----

The current chamber temperature is

 

 

 

 

 

 

 

 

displayed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press mode change key when

 

 

 

 

 

 

1

changing the set temperature

PV/SV

The second digits blinks.

 

 

 

 

 

during operation.

 

 

 

 

 

 

2

Set the second digit to 7 with the

 

When pressed, the figure of settable

numerical value shift key.

 

digit changes.

3

By pressing the digit shift key, set

 

The first digit blinks.

the first digit.

 

4

Set the first digit to 0 with the

 

When pressed, the figure of settable

numerical value shift key.

 

digit changes.

 

 

 

Set temperature is memorized and

5

Press enter key.

ENT

the current chamber temperature

 

 

 

 

 

 

 

is displayed.

 

 

 

 

 

 

 

 

 

 

 

Note: The temperature set mode returns to the temperature display mode automatically when 90 seconds has passed without any key operation.

Remote alarm terminal

The terminal of the remote alarm is located at rear of the machinery room, upper side of the electric box. The signal is contact output. The recommended contact capacity is between 2 A (DC 30V) .

Contact output: : normal close connect with N.O. and COM.

Consult our dealer or our office for installation of the remote alarm.

16

Page 17
Image 17
Sanyo MDF-793 Operating Instructions, Temperature setting, Remote alarm terminal, Description of operation, Key operated